Percentage of stenosis of a vascular lumen was quantitated using a digital subtraction angiography system with videodensitometric analysis. To validate the algorithm and methods, we used precisely drilled Lucite rods of three inner diameters and various reduced diameters to give known luminal reduction. Both in vitro and in vivo results of stenosis measurements resulted in an excellent correlation between actual and measured values (correlation coefficient > 0.9 for all trials). Consistent underestimation of stenosis of about 5% was attributed to the various image degradations inherent with the imaging procedure and equipment. Errors were greatest with mid range stenoses and less at the extremes of the lesion values. An insufficient signal-to-noise ratio was caused by low photon flux and/or low contrast material concentration and overlying vasculature filled with contrast material. Overall, the videodensitometric technique is accurate, easy to implement, objective, and relatively free of errors associated with the geometric/edge detection method.

Original languageEnglish (US)
Pages (from-to)807-811
Number of pages5
JournalRadiology
Volume157
Issue number3
StatePublished - 1985

Fingerprint

Pathologic Constriction
Contrast Media
Digital Subtraction Angiography
Signal-To-Noise Ratio
Polymethyl Methacrylate
Photons
Blood Vessels
Equipment and Supplies
In Vitro Techniques

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology

Cite this

Seibert, J. A., Link, D. P., Hines, H. H., & Baltaxe, H. A. (1985). Videodensitometric quantitation of stenosis: In vitro and in vivo validation. Radiology, 157(3), 807-811.

Videodensitometric quantitation of stenosis : In vitro and in vivo validation. / Seibert, J Anthony; Link, Daniel P; Hines, H. H.; Baltaxe, H. A.

In: Radiology, Vol. 157, No. 3, 1985, p. 807-811.

Research output: Contribution to journalArticle

Seibert, JA, Link, DP, Hines, HH & Baltaxe, HA 1985, 'Videodensitometric quantitation of stenosis: In vitro and in vivo validation', Radiology, vol. 157, no. 3, pp. 807-811.
Seibert, J Anthony ; Link, Daniel P ; Hines, H. H. ; Baltaxe, H. A. / Videodensitometric quantitation of stenosis : In vitro and in vivo validation. In: Radiology. 1985 ; Vol. 157, No. 3. pp. 807-811.
@article{edcf75a5f50b443f98edc4c837c8ecf1,
title = "Videodensitometric quantitation of stenosis: In vitro and in vivo validation",
abstract = "Percentage of stenosis of a vascular lumen was quantitated using a digital subtraction angiography system with videodensitometric analysis. To validate the algorithm and methods, we used precisely drilled Lucite rods of three inner diameters and various reduced diameters to give known luminal reduction. Both in vitro and in vivo results of stenosis measurements resulted in an excellent correlation between actual and measured values (correlation coefficient > 0.9 for all trials). Consistent underestimation of stenosis of about 5{\%} was attributed to the various image degradations inherent with the imaging procedure and equipment. Errors were greatest with mid range stenoses and less at the extremes of the lesion values. An insufficient signal-to-noise ratio was caused by low photon flux and/or low contrast material concentration and overlying vasculature filled with contrast material. Overall, the videodensitometric technique is accurate, easy to implement, objective, and relatively free of errors associated with the geometric/edge detection method.",
author = "Seibert, {J Anthony} and Link, {Daniel P} and Hines, {H. H.} and Baltaxe, {H. A.}",
year = "1985",
language = "English (US)",
volume = "157",
pages = "807--811",
journal = "Radiology",
issn = "0033-8419",
publisher = "Radiological Society of North America Inc.",
number = "3",

}

TY - JOUR

T1 - Videodensitometric quantitation of stenosis

T2 - In vitro and in vivo validation

AU - Seibert, J Anthony

AU - Link, Daniel P

AU - Hines, H. H.

AU - Baltaxe, H. A.

PY - 1985

Y1 - 1985

N2 - Percentage of stenosis of a vascular lumen was quantitated using a digital subtraction angiography system with videodensitometric analysis. To validate the algorithm and methods, we used precisely drilled Lucite rods of three inner diameters and various reduced diameters to give known luminal reduction. Both in vitro and in vivo results of stenosis measurements resulted in an excellent correlation between actual and measured values (correlation coefficient > 0.9 for all trials). Consistent underestimation of stenosis of about 5% was attributed to the various image degradations inherent with the imaging procedure and equipment. Errors were greatest with mid range stenoses and less at the extremes of the lesion values. An insufficient signal-to-noise ratio was caused by low photon flux and/or low contrast material concentration and overlying vasculature filled with contrast material. Overall, the videodensitometric technique is accurate, easy to implement, objective, and relatively free of errors associated with the geometric/edge detection method.

AB - Percentage of stenosis of a vascular lumen was quantitated using a digital subtraction angiography system with videodensitometric analysis. To validate the algorithm and methods, we used precisely drilled Lucite rods of three inner diameters and various reduced diameters to give known luminal reduction. Both in vitro and in vivo results of stenosis measurements resulted in an excellent correlation between actual and measured values (correlation coefficient > 0.9 for all trials). Consistent underestimation of stenosis of about 5% was attributed to the various image degradations inherent with the imaging procedure and equipment. Errors were greatest with mid range stenoses and less at the extremes of the lesion values. An insufficient signal-to-noise ratio was caused by low photon flux and/or low contrast material concentration and overlying vasculature filled with contrast material. Overall, the videodensitometric technique is accurate, easy to implement, objective, and relatively free of errors associated with the geometric/edge detection method.

UR - http://www.scopus.com/inward/record.url?scp=0022344117&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0022344117&partnerID=8YFLogxK

M3 - Article

C2 - 3903860

AN - SCOPUS:0022344117

VL - 157

SP - 807

EP - 811