Ultrahigh speed spectral / Fourier domain ophthalmic OCT imaging

Benjamin Potsaid, Iwona Gorczynska, Vivek Srinivasan, Yueli Chen, Jonathan Liu, James Jiang, Alex Cable, Jay S. Duker, James G. Fujimoto

Research output: Contribution to journalConference article

1 Citation (Scopus)

Abstract

Ultrahigh speed spectral / Fourier domain optical coherence tomography (OCT) using a CMOS line scan camera with image acquisition rates of 70,000-312,500 axial scans per second is investigated. Several design configurations are presented to illustrate design trade-offs between acquisition speed, sensitivity, resolution and sensitivity roll-off performance. We demonstrate: extended imaging range and improved sensitivity roll-off at 70,000 axial scans per second with 4096 camera pixels, high speed and ultrahigh resolution imaging at 106,382 axial scans per second, and ultrahigh speed imaging at 250,000-312,500 axial scans per second. Each configuration is characterized through optical testing and the trade-offs demonstrated with in vivo imaging of the fovea and optic disk in the human retina. We show dense homogeneous 3D-OCT volumetric data sets that were acquired by raster scanning at 250,000 axial scans per second, which is an order of magnitude faster than most current generation spectral / Fourier OCT instruments. OCT fundus images constructed from the 3D-OCT data have no noticeable discontinuity of retinal features and show that there are minimal motion artifacts. Using an improved sensitivity roll-off configuration at 70,000 axial scans per second, long cross sectional scans are acquired at high resolution for imaging large areas of the retina, including the fovea and optic disk. Using an ultrahigh speed configuration at 250,000 axial scans per second, the fine porous structures of the lamina cribrosa can be seen from slices extracted from a dense 3D data set. Rapid repeated imaging of a small volume (4D-OCT) enables time resolved visualization of the capillary network surrounding the INL and may show individual red blood cells. This capability could create the possibility for alternative techniques for quantifying capillary blood flow, which cannot be measured with Doppler OCT methods because of the capillary's perpendicular orientation to the optical beam. The results of this study suggest that high speed CMOS cameras can achieve a significant improvement in performance for ophthalmic imaging. This promises to have a powerful impact in clinical applications by improving early diagnosis, reproducibility of quantitative measurements and enabling more sensitive assessment of disease progression or response to therapy.

Original languageEnglish (US)
Article number716307
JournalProgress in Biomedical Optics and Imaging - Proceedings of SPIE
Volume7163
DOIs
StatePublished - May 13 2009
Externally publishedYes
EventOphthalmic Technologies XIX - San Jose, CA, United States
Duration: Jan 24 2009Jan 26 2009

Fingerprint

Optical tomography
Optical Coherence Tomography
tomography
Imaging techniques
fovea
retina
sensitivity
Cameras
cameras
Optic Disk
configurations
Retina
Optics
CMOS
acquisition
Blood
high speed
Optical testing
optics
raster scanning

Keywords

  • CMOS camera
  • Fovea and optic nerve head
  • Medical and biological imaging
  • Ophthalmology
  • Spectral / Fourier domain OCT
  • Ultrahigh speed optical coherence tomography

ASJC Scopus subject areas

  • Atomic and Molecular Physics, and Optics
  • Electronic, Optical and Magnetic Materials
  • Biomaterials
  • Radiology Nuclear Medicine and imaging

Cite this

Ultrahigh speed spectral / Fourier domain ophthalmic OCT imaging. / Potsaid, Benjamin; Gorczynska, Iwona; Srinivasan, Vivek; Chen, Yueli; Liu, Jonathan; Jiang, James; Cable, Alex; Duker, Jay S.; Fujimoto, James G.

In: Progress in Biomedical Optics and Imaging - Proceedings of SPIE, Vol. 7163, 716307, 13.05.2009.

Research output: Contribution to journalConference article

Potsaid, B, Gorczynska, I, Srinivasan, V, Chen, Y, Liu, J, Jiang, J, Cable, A, Duker, JS & Fujimoto, JG 2009, 'Ultrahigh speed spectral / Fourier domain ophthalmic OCT imaging', Progress in Biomedical Optics and Imaging - Proceedings of SPIE, vol. 7163, 716307. https://doi.org/10.1117/12.809058
Potsaid, Benjamin ; Gorczynska, Iwona ; Srinivasan, Vivek ; Chen, Yueli ; Liu, Jonathan ; Jiang, James ; Cable, Alex ; Duker, Jay S. ; Fujimoto, James G. / Ultrahigh speed spectral / Fourier domain ophthalmic OCT imaging. In: Progress in Biomedical Optics and Imaging - Proceedings of SPIE. 2009 ; Vol. 7163.
@article{804cd9798da74dfea67e57506516a1fc,
title = "Ultrahigh speed spectral / Fourier domain ophthalmic OCT imaging",
abstract = "Ultrahigh speed spectral / Fourier domain optical coherence tomography (OCT) using a CMOS line scan camera with image acquisition rates of 70,000-312,500 axial scans per second is investigated. Several design configurations are presented to illustrate design trade-offs between acquisition speed, sensitivity, resolution and sensitivity roll-off performance. We demonstrate: extended imaging range and improved sensitivity roll-off at 70,000 axial scans per second with 4096 camera pixels, high speed and ultrahigh resolution imaging at 106,382 axial scans per second, and ultrahigh speed imaging at 250,000-312,500 axial scans per second. Each configuration is characterized through optical testing and the trade-offs demonstrated with in vivo imaging of the fovea and optic disk in the human retina. We show dense homogeneous 3D-OCT volumetric data sets that were acquired by raster scanning at 250,000 axial scans per second, which is an order of magnitude faster than most current generation spectral / Fourier OCT instruments. OCT fundus images constructed from the 3D-OCT data have no noticeable discontinuity of retinal features and show that there are minimal motion artifacts. Using an improved sensitivity roll-off configuration at 70,000 axial scans per second, long cross sectional scans are acquired at high resolution for imaging large areas of the retina, including the fovea and optic disk. Using an ultrahigh speed configuration at 250,000 axial scans per second, the fine porous structures of the lamina cribrosa can be seen from slices extracted from a dense 3D data set. Rapid repeated imaging of a small volume (4D-OCT) enables time resolved visualization of the capillary network surrounding the INL and may show individual red blood cells. This capability could create the possibility for alternative techniques for quantifying capillary blood flow, which cannot be measured with Doppler OCT methods because of the capillary's perpendicular orientation to the optical beam. The results of this study suggest that high speed CMOS cameras can achieve a significant improvement in performance for ophthalmic imaging. This promises to have a powerful impact in clinical applications by improving early diagnosis, reproducibility of quantitative measurements and enabling more sensitive assessment of disease progression or response to therapy.",
keywords = "CMOS camera, Fovea and optic nerve head, Medical and biological imaging, Ophthalmology, Spectral / Fourier domain OCT, Ultrahigh speed optical coherence tomography",
author = "Benjamin Potsaid and Iwona Gorczynska and Vivek Srinivasan and Yueli Chen and Jonathan Liu and James Jiang and Alex Cable and Duker, {Jay S.} and Fujimoto, {James G.}",
year = "2009",
month = "5",
day = "13",
doi = "10.1117/12.809058",
language = "English (US)",
volume = "7163",
journal = "Progress in Biomedical Optics and Imaging - Proceedings of SPIE",
issn = "1605-7422",
publisher = "SPIE",

}

TY - JOUR

T1 - Ultrahigh speed spectral / Fourier domain ophthalmic OCT imaging

AU - Potsaid, Benjamin

AU - Gorczynska, Iwona

AU - Srinivasan, Vivek

AU - Chen, Yueli

AU - Liu, Jonathan

AU - Jiang, James

AU - Cable, Alex

AU - Duker, Jay S.

AU - Fujimoto, James G.

PY - 2009/5/13

Y1 - 2009/5/13

N2 - Ultrahigh speed spectral / Fourier domain optical coherence tomography (OCT) using a CMOS line scan camera with image acquisition rates of 70,000-312,500 axial scans per second is investigated. Several design configurations are presented to illustrate design trade-offs between acquisition speed, sensitivity, resolution and sensitivity roll-off performance. We demonstrate: extended imaging range and improved sensitivity roll-off at 70,000 axial scans per second with 4096 camera pixels, high speed and ultrahigh resolution imaging at 106,382 axial scans per second, and ultrahigh speed imaging at 250,000-312,500 axial scans per second. Each configuration is characterized through optical testing and the trade-offs demonstrated with in vivo imaging of the fovea and optic disk in the human retina. We show dense homogeneous 3D-OCT volumetric data sets that were acquired by raster scanning at 250,000 axial scans per second, which is an order of magnitude faster than most current generation spectral / Fourier OCT instruments. OCT fundus images constructed from the 3D-OCT data have no noticeable discontinuity of retinal features and show that there are minimal motion artifacts. Using an improved sensitivity roll-off configuration at 70,000 axial scans per second, long cross sectional scans are acquired at high resolution for imaging large areas of the retina, including the fovea and optic disk. Using an ultrahigh speed configuration at 250,000 axial scans per second, the fine porous structures of the lamina cribrosa can be seen from slices extracted from a dense 3D data set. Rapid repeated imaging of a small volume (4D-OCT) enables time resolved visualization of the capillary network surrounding the INL and may show individual red blood cells. This capability could create the possibility for alternative techniques for quantifying capillary blood flow, which cannot be measured with Doppler OCT methods because of the capillary's perpendicular orientation to the optical beam. The results of this study suggest that high speed CMOS cameras can achieve a significant improvement in performance for ophthalmic imaging. This promises to have a powerful impact in clinical applications by improving early diagnosis, reproducibility of quantitative measurements and enabling more sensitive assessment of disease progression or response to therapy.

AB - Ultrahigh speed spectral / Fourier domain optical coherence tomography (OCT) using a CMOS line scan camera with image acquisition rates of 70,000-312,500 axial scans per second is investigated. Several design configurations are presented to illustrate design trade-offs between acquisition speed, sensitivity, resolution and sensitivity roll-off performance. We demonstrate: extended imaging range and improved sensitivity roll-off at 70,000 axial scans per second with 4096 camera pixels, high speed and ultrahigh resolution imaging at 106,382 axial scans per second, and ultrahigh speed imaging at 250,000-312,500 axial scans per second. Each configuration is characterized through optical testing and the trade-offs demonstrated with in vivo imaging of the fovea and optic disk in the human retina. We show dense homogeneous 3D-OCT volumetric data sets that were acquired by raster scanning at 250,000 axial scans per second, which is an order of magnitude faster than most current generation spectral / Fourier OCT instruments. OCT fundus images constructed from the 3D-OCT data have no noticeable discontinuity of retinal features and show that there are minimal motion artifacts. Using an improved sensitivity roll-off configuration at 70,000 axial scans per second, long cross sectional scans are acquired at high resolution for imaging large areas of the retina, including the fovea and optic disk. Using an ultrahigh speed configuration at 250,000 axial scans per second, the fine porous structures of the lamina cribrosa can be seen from slices extracted from a dense 3D data set. Rapid repeated imaging of a small volume (4D-OCT) enables time resolved visualization of the capillary network surrounding the INL and may show individual red blood cells. This capability could create the possibility for alternative techniques for quantifying capillary blood flow, which cannot be measured with Doppler OCT methods because of the capillary's perpendicular orientation to the optical beam. The results of this study suggest that high speed CMOS cameras can achieve a significant improvement in performance for ophthalmic imaging. This promises to have a powerful impact in clinical applications by improving early diagnosis, reproducibility of quantitative measurements and enabling more sensitive assessment of disease progression or response to therapy.

KW - CMOS camera

KW - Fovea and optic nerve head

KW - Medical and biological imaging

KW - Ophthalmology

KW - Spectral / Fourier domain OCT

KW - Ultrahigh speed optical coherence tomography

UR - http://www.scopus.com/inward/record.url?scp=65549101671&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=65549101671&partnerID=8YFLogxK

U2 - 10.1117/12.809058

DO - 10.1117/12.809058

M3 - Conference article

AN - SCOPUS:65549101671

VL - 7163

JO - Progress in Biomedical Optics and Imaging - Proceedings of SPIE

JF - Progress in Biomedical Optics and Imaging - Proceedings of SPIE

SN - 1605-7422

M1 - 716307

ER -