Thoracic electric bioimpedance measurement of cardiac output in the newborn infant

Jaques Belik, Andrew N Pelech

Research output: Contribution to journalArticle

6 Citations (Scopus)

Abstract

To evaluate thoracic electric bioimpedance as a noninvasive method for measuring cardiac output, we compared the bioimpedance measurements with those obtained by means of the thermodilution indicator cardiac output technique in seven preterm and term lambs; we also studied 17 term and preterm infants. Sixty-seven simultaneous bioimpedance and thermodilution cardiac output measurements were obtained in the animals after intravascular volme expansion (saline solution infusion) and contraction (phlebotomy). A significant correlation between the cardiac output measurements by the two methods (0.82; p<0.001) was observed. In neonates, the observed cardiac output was 198±46 ml/kg·min in the preterm infants and 178±46 ml/kg ·min in the term infants. Extrapolating animal data to the neonates, we found the thoracic segment length recommended (the average of 29% body length and electrode distance) to be accurate. These data indicate that bioimpedance cardiac output measurement (1) is comparable to measurement by the thermodilution indicator technique in the newborn animal and (2) may be suitable for use infants and children.

Original languageEnglish (US)
Pages (from-to)890-895
Number of pages6
JournalThe Journal of Pediatrics
Volume113
Issue number5
DOIs
StatePublished - 1988
Externally publishedYes

Fingerprint

Cardiac Output
Thorax
Newborn Infant
Thermodilution
Premature Infants
Newborn Animals
Phlebotomy
Sodium Chloride
Electrodes

ASJC Scopus subject areas

  • Pediatrics, Perinatology, and Child Health
  • Medicine(all)

Cite this

Thoracic electric bioimpedance measurement of cardiac output in the newborn infant. / Belik, Jaques; Pelech, Andrew N.

In: The Journal of Pediatrics, Vol. 113, No. 5, 1988, p. 890-895.

Research output: Contribution to journalArticle

@article{3ea6fc2dd8484a6dbde4ea02c2893482,
title = "Thoracic electric bioimpedance measurement of cardiac output in the newborn infant",
abstract = "To evaluate thoracic electric bioimpedance as a noninvasive method for measuring cardiac output, we compared the bioimpedance measurements with those obtained by means of the thermodilution indicator cardiac output technique in seven preterm and term lambs; we also studied 17 term and preterm infants. Sixty-seven simultaneous bioimpedance and thermodilution cardiac output measurements were obtained in the animals after intravascular volme expansion (saline solution infusion) and contraction (phlebotomy). A significant correlation between the cardiac output measurements by the two methods (0.82; p<0.001) was observed. In neonates, the observed cardiac output was 198±46 ml/kg·min in the preterm infants and 178±46 ml/kg ·min in the term infants. Extrapolating animal data to the neonates, we found the thoracic segment length recommended (the average of 29{\%} body length and electrode distance) to be accurate. These data indicate that bioimpedance cardiac output measurement (1) is comparable to measurement by the thermodilution indicator technique in the newborn animal and (2) may be suitable for use infants and children.",
author = "Jaques Belik and Pelech, {Andrew N}",
year = "1988",
doi = "10.1016/S0022-3476(88)80026-X",
language = "English (US)",
volume = "113",
pages = "890--895",
journal = "Journal of Pediatrics",
issn = "0022-3476",
publisher = "Mosby Inc.",
number = "5",

}

TY - JOUR

T1 - Thoracic electric bioimpedance measurement of cardiac output in the newborn infant

AU - Belik, Jaques

AU - Pelech, Andrew N

PY - 1988

Y1 - 1988

N2 - To evaluate thoracic electric bioimpedance as a noninvasive method for measuring cardiac output, we compared the bioimpedance measurements with those obtained by means of the thermodilution indicator cardiac output technique in seven preterm and term lambs; we also studied 17 term and preterm infants. Sixty-seven simultaneous bioimpedance and thermodilution cardiac output measurements were obtained in the animals after intravascular volme expansion (saline solution infusion) and contraction (phlebotomy). A significant correlation between the cardiac output measurements by the two methods (0.82; p<0.001) was observed. In neonates, the observed cardiac output was 198±46 ml/kg·min in the preterm infants and 178±46 ml/kg ·min in the term infants. Extrapolating animal data to the neonates, we found the thoracic segment length recommended (the average of 29% body length and electrode distance) to be accurate. These data indicate that bioimpedance cardiac output measurement (1) is comparable to measurement by the thermodilution indicator technique in the newborn animal and (2) may be suitable for use infants and children.

AB - To evaluate thoracic electric bioimpedance as a noninvasive method for measuring cardiac output, we compared the bioimpedance measurements with those obtained by means of the thermodilution indicator cardiac output technique in seven preterm and term lambs; we also studied 17 term and preterm infants. Sixty-seven simultaneous bioimpedance and thermodilution cardiac output measurements were obtained in the animals after intravascular volme expansion (saline solution infusion) and contraction (phlebotomy). A significant correlation between the cardiac output measurements by the two methods (0.82; p<0.001) was observed. In neonates, the observed cardiac output was 198±46 ml/kg·min in the preterm infants and 178±46 ml/kg ·min in the term infants. Extrapolating animal data to the neonates, we found the thoracic segment length recommended (the average of 29% body length and electrode distance) to be accurate. These data indicate that bioimpedance cardiac output measurement (1) is comparable to measurement by the thermodilution indicator technique in the newborn animal and (2) may be suitable for use infants and children.

UR - http://www.scopus.com/inward/record.url?scp=0023701194&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0023701194&partnerID=8YFLogxK

U2 - 10.1016/S0022-3476(88)80026-X

DO - 10.1016/S0022-3476(88)80026-X

M3 - Article

C2 - 3183850

AN - SCOPUS:0023701194

VL - 113

SP - 890

EP - 895

JO - Journal of Pediatrics

JF - Journal of Pediatrics

SN - 0022-3476

IS - 5

ER -