The effect of skin thickness determined using breast CT on mammographic dosimetry

Shih Ying Huang, John M Boone, Kai Yang, Alexander L C Kwan, Nathan J. Packard

Research output: Contribution to journalArticle

73 Citations (Scopus)

Abstract

The effect of breast skin thickness on dosimetry in mammography was investigated. Breast computed tomography (CT) acquisition techniques, combined with algorithms designed for determining specific breast metrics, were useful for estimating skin thickness. A radial-geometry edge detection scheme was implemented on coronal reconstructed breast CT (bCT) images to measure the breast skin thickness. Skin thickness of bilateral bCT volume data from 49 women and unilateral bCT volume data from 2 women (10 healthy women and 41 women with BIRADS 4 and 5 diagnoses) was robustly measured with the edge detection scheme. The mean breast skin thickness (±inter -breast standard deviation) was found to be 1.45±0.30 mm. Since most current published normalized glandular dose (DgN) coefficients are based on the assumption of a 4-mm breast skin thickness, the DgN values computed with Monte Carlo techniques will increase up to 18% due to the thinner skin layers (e.g., 6-cm 50% glandular breast, 28 kVp Mo-Mo spectrum). The thinner skin dimensions found in this study suggest that the current DgN values used for mammographic dosimetry lead to a slight underestimate in glandular dose.

Original languageEnglish (US)
Pages (from-to)1199-1206
Number of pages8
JournalMedical Physics
Volume35
Issue number4
DOIs
StatePublished - 2008

Fingerprint

Breast
Tomography
Skin
Cone-Beam Computed Tomography
Mammography

Keywords

  • Breast CT
  • Breast skin thickness
  • Edge detection
  • Normalized glandular dose coefficients

ASJC Scopus subject areas

  • Biophysics

Cite this

The effect of skin thickness determined using breast CT on mammographic dosimetry. / Huang, Shih Ying; Boone, John M; Yang, Kai; Kwan, Alexander L C; Packard, Nathan J.

In: Medical Physics, Vol. 35, No. 4, 2008, p. 1199-1206.

Research output: Contribution to journalArticle

Huang, Shih Ying ; Boone, John M ; Yang, Kai ; Kwan, Alexander L C ; Packard, Nathan J. / The effect of skin thickness determined using breast CT on mammographic dosimetry. In: Medical Physics. 2008 ; Vol. 35, No. 4. pp. 1199-1206.
@article{42bd358200f643379d76226260d5139c,
title = "The effect of skin thickness determined using breast CT on mammographic dosimetry",
abstract = "The effect of breast skin thickness on dosimetry in mammography was investigated. Breast computed tomography (CT) acquisition techniques, combined with algorithms designed for determining specific breast metrics, were useful for estimating skin thickness. A radial-geometry edge detection scheme was implemented on coronal reconstructed breast CT (bCT) images to measure the breast skin thickness. Skin thickness of bilateral bCT volume data from 49 women and unilateral bCT volume data from 2 women (10 healthy women and 41 women with BIRADS 4 and 5 diagnoses) was robustly measured with the edge detection scheme. The mean breast skin thickness (±inter -breast standard deviation) was found to be 1.45±0.30 mm. Since most current published normalized glandular dose (DgN) coefficients are based on the assumption of a 4-mm breast skin thickness, the DgN values computed with Monte Carlo techniques will increase up to 18{\%} due to the thinner skin layers (e.g., 6-cm 50{\%} glandular breast, 28 kVp Mo-Mo spectrum). The thinner skin dimensions found in this study suggest that the current DgN values used for mammographic dosimetry lead to a slight underestimate in glandular dose.",
keywords = "Breast CT, Breast skin thickness, Edge detection, Normalized glandular dose coefficients",
author = "Huang, {Shih Ying} and Boone, {John M} and Kai Yang and Kwan, {Alexander L C} and Packard, {Nathan J.}",
year = "2008",
doi = "10.1118/1.2841938",
language = "English (US)",
volume = "35",
pages = "1199--1206",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "4",

}

TY - JOUR

T1 - The effect of skin thickness determined using breast CT on mammographic dosimetry

AU - Huang, Shih Ying

AU - Boone, John M

AU - Yang, Kai

AU - Kwan, Alexander L C

AU - Packard, Nathan J.

PY - 2008

Y1 - 2008

N2 - The effect of breast skin thickness on dosimetry in mammography was investigated. Breast computed tomography (CT) acquisition techniques, combined with algorithms designed for determining specific breast metrics, were useful for estimating skin thickness. A radial-geometry edge detection scheme was implemented on coronal reconstructed breast CT (bCT) images to measure the breast skin thickness. Skin thickness of bilateral bCT volume data from 49 women and unilateral bCT volume data from 2 women (10 healthy women and 41 women with BIRADS 4 and 5 diagnoses) was robustly measured with the edge detection scheme. The mean breast skin thickness (±inter -breast standard deviation) was found to be 1.45±0.30 mm. Since most current published normalized glandular dose (DgN) coefficients are based on the assumption of a 4-mm breast skin thickness, the DgN values computed with Monte Carlo techniques will increase up to 18% due to the thinner skin layers (e.g., 6-cm 50% glandular breast, 28 kVp Mo-Mo spectrum). The thinner skin dimensions found in this study suggest that the current DgN values used for mammographic dosimetry lead to a slight underestimate in glandular dose.

AB - The effect of breast skin thickness on dosimetry in mammography was investigated. Breast computed tomography (CT) acquisition techniques, combined with algorithms designed for determining specific breast metrics, were useful for estimating skin thickness. A radial-geometry edge detection scheme was implemented on coronal reconstructed breast CT (bCT) images to measure the breast skin thickness. Skin thickness of bilateral bCT volume data from 49 women and unilateral bCT volume data from 2 women (10 healthy women and 41 women with BIRADS 4 and 5 diagnoses) was robustly measured with the edge detection scheme. The mean breast skin thickness (±inter -breast standard deviation) was found to be 1.45±0.30 mm. Since most current published normalized glandular dose (DgN) coefficients are based on the assumption of a 4-mm breast skin thickness, the DgN values computed with Monte Carlo techniques will increase up to 18% due to the thinner skin layers (e.g., 6-cm 50% glandular breast, 28 kVp Mo-Mo spectrum). The thinner skin dimensions found in this study suggest that the current DgN values used for mammographic dosimetry lead to a slight underestimate in glandular dose.

KW - Breast CT

KW - Breast skin thickness

KW - Edge detection

KW - Normalized glandular dose coefficients

UR - http://www.scopus.com/inward/record.url?scp=41449092092&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=41449092092&partnerID=8YFLogxK

U2 - 10.1118/1.2841938

DO - 10.1118/1.2841938

M3 - Article

C2 - 18491511

AN - SCOPUS:41449092092

VL - 35

SP - 1199

EP - 1206

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 4

ER -