Commissioning and QA of Cone Beam CT for Image Guided Radiation Therapy

J. Lehmann, Julian R Perks, S. Semon, J. Purdy

Research output: Contribution to journalArticle

Abstract

Purpose: To report the results of commissioning tests and the development of a quality assurance program for the commercial version (i.e., non‐research model) Elekta Synergy Cone Beam CT (CBCT) system. Methods and Materials: Our institution was one of the first sites in the US to install a commercial version of the Elekta Synergy Cone Beam CT system. A commissioning procedure was developed consisting of six parts: (1) System safety, (2) Geometric accuracy (i.e., agreement of MV and kV beam isocenters), (3) Image quality, (4) Algorithm accuracy, (5) Dose to patient, and (6) Quality assurance procedures. Results: The system passed all safety tests and agreement of MV/kV isocenters was found to be within 1 mm. Using a precisely moved skull phantom, reconstruction and alignment algorithm was found to be accurate within 1mm and 1 degree in each dimension. For 12 measurement points spanning a 9×9×15cm3 volume in a RANDO phantom, average agreement in x, y, and z coordinates was 0.010cm, −0.012cm, 0.022cm (SD: 0.021cm, 0.055cm, 0.021cm; Largest deviations: 0.06cm, 0.10cm, 0.05cm, respectively). Larger deviation for the y‐component can be partially attributed to CT slice thickness of 1 mm along that direction. Dose to patient depends on machine settings and was found not to exceed 6.3cGy in the highest setting used in our clinic (120kVp, 40mA, 40ms per frame, 360 degree scan, S20 field of view). A more typical dose for a full 360 degree scan is 4cGy. Conclusions: The procedures discussed in this presentation were applied to the Elekta Synergy CBCT system. They have been useful in identifying difficulties with the CBCT operation operation, including data transfer issues. System safety, geometric accuracy, and image quality checks are recommended to be performed on a monthly basis; algorithm accuracy and dose to patient checks are recommended to be performed annually.

Original languageEnglish (US)
Pages (from-to)2284-2285
Number of pages2
JournalMedical Physics
Volume33
Issue number6
DOIs
StatePublished - 2006

Fingerprint

Image-Guided Radiotherapy
Cone-Beam Computed Tomography
Safety
Skull

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

TH‐D‐224C‐10 : Commissioning and QA of Cone Beam CT for Image Guided Radiation Therapy. / Lehmann, J.; Perks, Julian R; Semon, S.; Purdy, J.

In: Medical Physics, Vol. 33, No. 6, 2006, p. 2284-2285.

Research output: Contribution to journalArticle

@article{85ed78a3bed84b9e86a43b96ab443564,
title = "TH‐D‐224C‐10: Commissioning and QA of Cone Beam CT for Image Guided Radiation Therapy",
abstract = "Purpose: To report the results of commissioning tests and the development of a quality assurance program for the commercial version (i.e., non‐research model) Elekta Synergy Cone Beam CT (CBCT) system. Methods and Materials: Our institution was one of the first sites in the US to install a commercial version of the Elekta Synergy Cone Beam CT system. A commissioning procedure was developed consisting of six parts: (1) System safety, (2) Geometric accuracy (i.e., agreement of MV and kV beam isocenters), (3) Image quality, (4) Algorithm accuracy, (5) Dose to patient, and (6) Quality assurance procedures. Results: The system passed all safety tests and agreement of MV/kV isocenters was found to be within 1 mm. Using a precisely moved skull phantom, reconstruction and alignment algorithm was found to be accurate within 1mm and 1 degree in each dimension. For 12 measurement points spanning a 9×9×15cm3 volume in a RANDO phantom, average agreement in x, y, and z coordinates was 0.010cm, −0.012cm, 0.022cm (SD: 0.021cm, 0.055cm, 0.021cm; Largest deviations: 0.06cm, 0.10cm, 0.05cm, respectively). Larger deviation for the y‐component can be partially attributed to CT slice thickness of 1 mm along that direction. Dose to patient depends on machine settings and was found not to exceed 6.3cGy in the highest setting used in our clinic (120kVp, 40mA, 40ms per frame, 360 degree scan, S20 field of view). A more typical dose for a full 360 degree scan is 4cGy. Conclusions: The procedures discussed in this presentation were applied to the Elekta Synergy CBCT system. They have been useful in identifying difficulties with the CBCT operation operation, including data transfer issues. System safety, geometric accuracy, and image quality checks are recommended to be performed on a monthly basis; algorithm accuracy and dose to patient checks are recommended to be performed annually.",
author = "J. Lehmann and Perks, {Julian R} and S. Semon and J. Purdy",
year = "2006",
doi = "10.1118/1.2241922",
language = "English (US)",
volume = "33",
pages = "2284--2285",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "6",

}

TY - JOUR

T1 - TH‐D‐224C‐10

T2 - Commissioning and QA of Cone Beam CT for Image Guided Radiation Therapy

AU - Lehmann, J.

AU - Perks, Julian R

AU - Semon, S.

AU - Purdy, J.

PY - 2006

Y1 - 2006

N2 - Purpose: To report the results of commissioning tests and the development of a quality assurance program for the commercial version (i.e., non‐research model) Elekta Synergy Cone Beam CT (CBCT) system. Methods and Materials: Our institution was one of the first sites in the US to install a commercial version of the Elekta Synergy Cone Beam CT system. A commissioning procedure was developed consisting of six parts: (1) System safety, (2) Geometric accuracy (i.e., agreement of MV and kV beam isocenters), (3) Image quality, (4) Algorithm accuracy, (5) Dose to patient, and (6) Quality assurance procedures. Results: The system passed all safety tests and agreement of MV/kV isocenters was found to be within 1 mm. Using a precisely moved skull phantom, reconstruction and alignment algorithm was found to be accurate within 1mm and 1 degree in each dimension. For 12 measurement points spanning a 9×9×15cm3 volume in a RANDO phantom, average agreement in x, y, and z coordinates was 0.010cm, −0.012cm, 0.022cm (SD: 0.021cm, 0.055cm, 0.021cm; Largest deviations: 0.06cm, 0.10cm, 0.05cm, respectively). Larger deviation for the y‐component can be partially attributed to CT slice thickness of 1 mm along that direction. Dose to patient depends on machine settings and was found not to exceed 6.3cGy in the highest setting used in our clinic (120kVp, 40mA, 40ms per frame, 360 degree scan, S20 field of view). A more typical dose for a full 360 degree scan is 4cGy. Conclusions: The procedures discussed in this presentation were applied to the Elekta Synergy CBCT system. They have been useful in identifying difficulties with the CBCT operation operation, including data transfer issues. System safety, geometric accuracy, and image quality checks are recommended to be performed on a monthly basis; algorithm accuracy and dose to patient checks are recommended to be performed annually.

AB - Purpose: To report the results of commissioning tests and the development of a quality assurance program for the commercial version (i.e., non‐research model) Elekta Synergy Cone Beam CT (CBCT) system. Methods and Materials: Our institution was one of the first sites in the US to install a commercial version of the Elekta Synergy Cone Beam CT system. A commissioning procedure was developed consisting of six parts: (1) System safety, (2) Geometric accuracy (i.e., agreement of MV and kV beam isocenters), (3) Image quality, (4) Algorithm accuracy, (5) Dose to patient, and (6) Quality assurance procedures. Results: The system passed all safety tests and agreement of MV/kV isocenters was found to be within 1 mm. Using a precisely moved skull phantom, reconstruction and alignment algorithm was found to be accurate within 1mm and 1 degree in each dimension. For 12 measurement points spanning a 9×9×15cm3 volume in a RANDO phantom, average agreement in x, y, and z coordinates was 0.010cm, −0.012cm, 0.022cm (SD: 0.021cm, 0.055cm, 0.021cm; Largest deviations: 0.06cm, 0.10cm, 0.05cm, respectively). Larger deviation for the y‐component can be partially attributed to CT slice thickness of 1 mm along that direction. Dose to patient depends on machine settings and was found not to exceed 6.3cGy in the highest setting used in our clinic (120kVp, 40mA, 40ms per frame, 360 degree scan, S20 field of view). A more typical dose for a full 360 degree scan is 4cGy. Conclusions: The procedures discussed in this presentation were applied to the Elekta Synergy CBCT system. They have been useful in identifying difficulties with the CBCT operation operation, including data transfer issues. System safety, geometric accuracy, and image quality checks are recommended to be performed on a monthly basis; algorithm accuracy and dose to patient checks are recommended to be performed annually.

UR - http://www.scopus.com/inward/record.url?scp=85024805597&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85024805597&partnerID=8YFLogxK

U2 - 10.1118/1.2241922

DO - 10.1118/1.2241922

M3 - Article

VL - 33

SP - 2284

EP - 2285

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 6

ER -