Purpose: To investigate the statistical reproducibility of probability distribution function (PDF) of intra‐fractional lung motion using dynamic magnetic resonance imaging (MRI). Method and Materials: 9 healthy subjects and 3 lung tumor patients underwent two continuous 5‐minutes MRI scans in the sagittal plane, repeated 2 weeks apart. Three pulmonary vessels from different lung regions (upper, middle, and lower) in healthy subjects and the lung tumors in patients were selected for tracking, and the displacement PDF reproducibility was evaluated as a function of scan time and frame rate. Results: In all 9 healthy subjects, the PDF reproducibility errors at 5‐minutes (mean 0.15, range 0.05 – 0.30) were significantly (p<0.0001) reduced compared to those at 5‐seconds (mean 0.41, range 0.25 – 0.70). In 3 patients, the PDF reproducibility errors at 5‐minutes (0.19, 0.10, 0.09 respectively) were also largely reduced from those at 5‐second (0.96, 0.41, and 0.24 respectively). PDF stabilizing time constants showed great variability among subjects (range 8 s – 374 s). 8 of 9 healthy volunteers and all 3 lung tumor patients were observed to reach PDF equilibrium state within 5 minutes. PDF reproducibility showed less sensitivity to frame rate when it is greater than 2 fps. Conclusions: A statistically significant improvement in the PDF reproducibility was observed with prolonged scan time among 12 healthy volunteers and patients with lung tumors. The improved knowledge of PDF reproducibility is a vital part of the validation of probability based inverse treatment planning for stereotactic radiosurgery of lung cancer.

Original languageEnglish (US)
Number of pages1
JournalMedical Physics
Volume34
Issue number6
DOIs
StatePublished - 2007
Externally publishedYes

Fingerprint

Magnetic Resonance Imaging
Lung
Healthy Volunteers
Neoplasms
Radiosurgery
Lung Neoplasms

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

@article{522c4ecd4b7a4f2aa848d0d964a37e7b,
title = "TH‐C‐L100J‐04: Reproducibility of Intra‐Fractional Lung Motion Probability Distribution Function (PDF) Using Dynamic MRI: Statistical Analysis",
abstract = "Purpose: To investigate the statistical reproducibility of probability distribution function (PDF) of intra‐fractional lung motion using dynamic magnetic resonance imaging (MRI). Method and Materials: 9 healthy subjects and 3 lung tumor patients underwent two continuous 5‐minutes MRI scans in the sagittal plane, repeated 2 weeks apart. Three pulmonary vessels from different lung regions (upper, middle, and lower) in healthy subjects and the lung tumors in patients were selected for tracking, and the displacement PDF reproducibility was evaluated as a function of scan time and frame rate. Results: In all 9 healthy subjects, the PDF reproducibility errors at 5‐minutes (mean 0.15, range 0.05 – 0.30) were significantly (p<0.0001) reduced compared to those at 5‐seconds (mean 0.41, range 0.25 – 0.70). In 3 patients, the PDF reproducibility errors at 5‐minutes (0.19, 0.10, 0.09 respectively) were also largely reduced from those at 5‐second (0.96, 0.41, and 0.24 respectively). PDF stabilizing time constants showed great variability among subjects (range 8 s – 374 s). 8 of 9 healthy volunteers and all 3 lung tumor patients were observed to reach PDF equilibrium state within 5 minutes. PDF reproducibility showed less sensitivity to frame rate when it is greater than 2 fps. Conclusions: A statistically significant improvement in the PDF reproducibility was observed with prolonged scan time among 12 healthy volunteers and patients with lung tumors. The improved knowledge of PDF reproducibility is a vital part of the validation of probability based inverse treatment planning for stereotactic radiosurgery of lung cancer.",
author = "J. Cai and P. Read and Benedict, {Stanley H} and J. Larner and K. Sheng",
year = "2007",
doi = "10.1118/1.2761642",
language = "English (US)",
volume = "34",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "6",

}

TY - JOUR

T1 - TH‐C‐L100J‐04

T2 - Reproducibility of Intra‐Fractional Lung Motion Probability Distribution Function (PDF) Using Dynamic MRI: Statistical Analysis

AU - Cai, J.

AU - Read, P.

AU - Benedict, Stanley H

AU - Larner, J.

AU - Sheng, K.

PY - 2007

Y1 - 2007

N2 - Purpose: To investigate the statistical reproducibility of probability distribution function (PDF) of intra‐fractional lung motion using dynamic magnetic resonance imaging (MRI). Method and Materials: 9 healthy subjects and 3 lung tumor patients underwent two continuous 5‐minutes MRI scans in the sagittal plane, repeated 2 weeks apart. Three pulmonary vessels from different lung regions (upper, middle, and lower) in healthy subjects and the lung tumors in patients were selected for tracking, and the displacement PDF reproducibility was evaluated as a function of scan time and frame rate. Results: In all 9 healthy subjects, the PDF reproducibility errors at 5‐minutes (mean 0.15, range 0.05 – 0.30) were significantly (p<0.0001) reduced compared to those at 5‐seconds (mean 0.41, range 0.25 – 0.70). In 3 patients, the PDF reproducibility errors at 5‐minutes (0.19, 0.10, 0.09 respectively) were also largely reduced from those at 5‐second (0.96, 0.41, and 0.24 respectively). PDF stabilizing time constants showed great variability among subjects (range 8 s – 374 s). 8 of 9 healthy volunteers and all 3 lung tumor patients were observed to reach PDF equilibrium state within 5 minutes. PDF reproducibility showed less sensitivity to frame rate when it is greater than 2 fps. Conclusions: A statistically significant improvement in the PDF reproducibility was observed with prolonged scan time among 12 healthy volunteers and patients with lung tumors. The improved knowledge of PDF reproducibility is a vital part of the validation of probability based inverse treatment planning for stereotactic radiosurgery of lung cancer.

AB - Purpose: To investigate the statistical reproducibility of probability distribution function (PDF) of intra‐fractional lung motion using dynamic magnetic resonance imaging (MRI). Method and Materials: 9 healthy subjects and 3 lung tumor patients underwent two continuous 5‐minutes MRI scans in the sagittal plane, repeated 2 weeks apart. Three pulmonary vessels from different lung regions (upper, middle, and lower) in healthy subjects and the lung tumors in patients were selected for tracking, and the displacement PDF reproducibility was evaluated as a function of scan time and frame rate. Results: In all 9 healthy subjects, the PDF reproducibility errors at 5‐minutes (mean 0.15, range 0.05 – 0.30) were significantly (p<0.0001) reduced compared to those at 5‐seconds (mean 0.41, range 0.25 – 0.70). In 3 patients, the PDF reproducibility errors at 5‐minutes (0.19, 0.10, 0.09 respectively) were also largely reduced from those at 5‐second (0.96, 0.41, and 0.24 respectively). PDF stabilizing time constants showed great variability among subjects (range 8 s – 374 s). 8 of 9 healthy volunteers and all 3 lung tumor patients were observed to reach PDF equilibrium state within 5 minutes. PDF reproducibility showed less sensitivity to frame rate when it is greater than 2 fps. Conclusions: A statistically significant improvement in the PDF reproducibility was observed with prolonged scan time among 12 healthy volunteers and patients with lung tumors. The improved knowledge of PDF reproducibility is a vital part of the validation of probability based inverse treatment planning for stereotactic radiosurgery of lung cancer.

UR - http://www.scopus.com/inward/record.url?scp=85024813167&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85024813167&partnerID=8YFLogxK

U2 - 10.1118/1.2761642

DO - 10.1118/1.2761642

M3 - Article

VL - 34

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 6

ER -