TH‐C‐AUD A‐04: Calibration of Image‐Guided Robotic System for Prostate Brachytherapy

I. Buzurovic, T. Podder, Y. hu, K. Yan, Richard K Valicenti, A. Dicker, Y. yu

Research output: Contribution to journalArticle

Abstract

Purpose: To perform calibration and evaluation of an image‐guided brachytherapy robotic system. Method and Materials: Brachytherapy robot calibration consists of three steps: mechanical robotic calibration, image calibration and overall system calibration. Mechanical calibration determines 1)system resolution, as the smallest incremental movement that the robot can physically perform, 2)repeatability as a measure of the ability of the robot to move back to the same position and orientation and 3)accuracy, as the robot's ability to precisely move to a desired position in 3D space. Imaging calibration for our system is the procedure where distance from ultrasound images is transformed to the metric distance in robot absolute coordinates. Overall system calibration is to determine the exact position of image stacks, calculated in absolute robotic system coordinates. These steps were performed using high‐resolution camera, specially‐designed phantom box for imaging calibration and CMM device for mechanical calibration. The system software allows mutual calibration between mechanical and imaging robotic modules. Mechanical calibration consists of DHV table definition for robotic system, matrix transformation, definition of composite matrices, direct kinematics solution, inverse kinematics solution, definition of robot initial position, calculation of position error, and error correction method. The robot errors gathered by position measurement are minimized by numerical optimization. Results: The calibrated precision of translation movements for the stylet and cannula are in the range of 0.03–0.08mm (depending on load); lateral and vertical movements for the gantry are 0.03mm; probe translation and rotation are 0.05mm and 0.03deg, respectively. The fiducial error for imaging is less than 0.1mm in x and y image coordinates. Conclusion: Overall robotic brachytherapy calibration plays a crucial role in accurate delivery. The calibrated precisions of the image‐guided brachytherapy robotic system are considered satisfactory for the given clinical application. Acknowledgement: Supported by NCI‐R01‐CA091763.

Original languageEnglish (US)
Pages (from-to)2970
Number of pages1
JournalMedical Physics
Volume35
Issue number6
DOIs
StatePublished - 2008
Externally publishedYes

Fingerprint

Brachytherapy
Robotics
Calibration
Prostate
Biomechanical Phenomena
Imaging Phantoms
Software

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

TH‐C‐AUD A‐04 : Calibration of Image‐Guided Robotic System for Prostate Brachytherapy. / Buzurovic, I.; Podder, T.; hu, Y.; Yan, K.; Valicenti, Richard K; Dicker, A.; yu, Y.

In: Medical Physics, Vol. 35, No. 6, 2008, p. 2970.

Research output: Contribution to journalArticle

Buzurovic, I. ; Podder, T. ; hu, Y. ; Yan, K. ; Valicenti, Richard K ; Dicker, A. ; yu, Y. / TH‐C‐AUD A‐04 : Calibration of Image‐Guided Robotic System for Prostate Brachytherapy. In: Medical Physics. 2008 ; Vol. 35, No. 6. pp. 2970.
@article{0cca3601ffe4473a9a0d3352bf62b3ae,
title = "TH‐C‐AUD A‐04: Calibration of Image‐Guided Robotic System for Prostate Brachytherapy",
abstract = "Purpose: To perform calibration and evaluation of an image‐guided brachytherapy robotic system. Method and Materials: Brachytherapy robot calibration consists of three steps: mechanical robotic calibration, image calibration and overall system calibration. Mechanical calibration determines 1)system resolution, as the smallest incremental movement that the robot can physically perform, 2)repeatability as a measure of the ability of the robot to move back to the same position and orientation and 3)accuracy, as the robot's ability to precisely move to a desired position in 3D space. Imaging calibration for our system is the procedure where distance from ultrasound images is transformed to the metric distance in robot absolute coordinates. Overall system calibration is to determine the exact position of image stacks, calculated in absolute robotic system coordinates. These steps were performed using high‐resolution camera, specially‐designed phantom box for imaging calibration and CMM device for mechanical calibration. The system software allows mutual calibration between mechanical and imaging robotic modules. Mechanical calibration consists of DHV table definition for robotic system, matrix transformation, definition of composite matrices, direct kinematics solution, inverse kinematics solution, definition of robot initial position, calculation of position error, and error correction method. The robot errors gathered by position measurement are minimized by numerical optimization. Results: The calibrated precision of translation movements for the stylet and cannula are in the range of 0.03–0.08mm (depending on load); lateral and vertical movements for the gantry are 0.03mm; probe translation and rotation are 0.05mm and 0.03deg, respectively. The fiducial error for imaging is less than 0.1mm in x and y image coordinates. Conclusion: Overall robotic brachytherapy calibration plays a crucial role in accurate delivery. The calibrated precisions of the image‐guided brachytherapy robotic system are considered satisfactory for the given clinical application. Acknowledgement: Supported by NCI‐R01‐CA091763.",
author = "I. Buzurovic and T. Podder and Y. hu and K. Yan and Valicenti, {Richard K} and A. Dicker and Y. yu",
year = "2008",
doi = "10.1118/1.2962839",
language = "English (US)",
volume = "35",
pages = "2970",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "6",

}

TY - JOUR

T1 - TH‐C‐AUD A‐04

T2 - Calibration of Image‐Guided Robotic System for Prostate Brachytherapy

AU - Buzurovic, I.

AU - Podder, T.

AU - hu, Y.

AU - Yan, K.

AU - Valicenti, Richard K

AU - Dicker, A.

AU - yu, Y.

PY - 2008

Y1 - 2008

N2 - Purpose: To perform calibration and evaluation of an image‐guided brachytherapy robotic system. Method and Materials: Brachytherapy robot calibration consists of three steps: mechanical robotic calibration, image calibration and overall system calibration. Mechanical calibration determines 1)system resolution, as the smallest incremental movement that the robot can physically perform, 2)repeatability as a measure of the ability of the robot to move back to the same position and orientation and 3)accuracy, as the robot's ability to precisely move to a desired position in 3D space. Imaging calibration for our system is the procedure where distance from ultrasound images is transformed to the metric distance in robot absolute coordinates. Overall system calibration is to determine the exact position of image stacks, calculated in absolute robotic system coordinates. These steps were performed using high‐resolution camera, specially‐designed phantom box for imaging calibration and CMM device for mechanical calibration. The system software allows mutual calibration between mechanical and imaging robotic modules. Mechanical calibration consists of DHV table definition for robotic system, matrix transformation, definition of composite matrices, direct kinematics solution, inverse kinematics solution, definition of robot initial position, calculation of position error, and error correction method. The robot errors gathered by position measurement are minimized by numerical optimization. Results: The calibrated precision of translation movements for the stylet and cannula are in the range of 0.03–0.08mm (depending on load); lateral and vertical movements for the gantry are 0.03mm; probe translation and rotation are 0.05mm and 0.03deg, respectively. The fiducial error for imaging is less than 0.1mm in x and y image coordinates. Conclusion: Overall robotic brachytherapy calibration plays a crucial role in accurate delivery. The calibrated precisions of the image‐guided brachytherapy robotic system are considered satisfactory for the given clinical application. Acknowledgement: Supported by NCI‐R01‐CA091763.

AB - Purpose: To perform calibration and evaluation of an image‐guided brachytherapy robotic system. Method and Materials: Brachytherapy robot calibration consists of three steps: mechanical robotic calibration, image calibration and overall system calibration. Mechanical calibration determines 1)system resolution, as the smallest incremental movement that the robot can physically perform, 2)repeatability as a measure of the ability of the robot to move back to the same position and orientation and 3)accuracy, as the robot's ability to precisely move to a desired position in 3D space. Imaging calibration for our system is the procedure where distance from ultrasound images is transformed to the metric distance in robot absolute coordinates. Overall system calibration is to determine the exact position of image stacks, calculated in absolute robotic system coordinates. These steps were performed using high‐resolution camera, specially‐designed phantom box for imaging calibration and CMM device for mechanical calibration. The system software allows mutual calibration between mechanical and imaging robotic modules. Mechanical calibration consists of DHV table definition for robotic system, matrix transformation, definition of composite matrices, direct kinematics solution, inverse kinematics solution, definition of robot initial position, calculation of position error, and error correction method. The robot errors gathered by position measurement are minimized by numerical optimization. Results: The calibrated precision of translation movements for the stylet and cannula are in the range of 0.03–0.08mm (depending on load); lateral and vertical movements for the gantry are 0.03mm; probe translation and rotation are 0.05mm and 0.03deg, respectively. The fiducial error for imaging is less than 0.1mm in x and y image coordinates. Conclusion: Overall robotic brachytherapy calibration plays a crucial role in accurate delivery. The calibrated precisions of the image‐guided brachytherapy robotic system are considered satisfactory for the given clinical application. Acknowledgement: Supported by NCI‐R01‐CA091763.

UR - http://www.scopus.com/inward/record.url?scp=85024786576&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85024786576&partnerID=8YFLogxK

U2 - 10.1118/1.2962839

DO - 10.1118/1.2962839

M3 - Article

AN - SCOPUS:85024786576

VL - 35

SP - 2970

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 6

ER -