Technique to control pH in vicinity of biodegrading PLA-PGA implants

C. Mauli Agrawal, Kyriacos A. Athanasiou

Research output: Contribution to journalArticle

368 Scopus citations

Abstract

This in vitro study was performed to examine if the pH decrease in the vicinity of degrading polylactic acid (PLA) and polyglycolic acid (PGA) polymers can be offset by incorporation of basic salts within PLA-PGA implants. It has been suggested that such pH lowering results in adverse effects, which may be responsible for biocompatibility concerns raised recently about PLA and PGA polymers. The results indicated that all three salts investigated in this study were successful in controlling the decrease in pH due to the acidic degradation products of the copolymer. The pH of the test media for the control group fell to a value of 3.0 at 9 weeks. Implants containing calcium carbonate maintained the pH value between 7.4 and 6.3 throughout the degradation process. Implants with calcium hydroxyapatite and sodium bicarbonate controlled the pH values between 6.9 and 4.3 and 8.2 and 4.5, respectively. At 3 weeks, marked swelling of implants containing calcium carbonate or sodium bicarbonate was observed relative to the control implants. The molecular weight and mass changes in the implants did not show any significant differences at 9 weeks. Thus, results from this in vitro model show that a significant decrease in pH in the vicinity of PLA-PGA implants can be avoided by incorporating basic salts.

Original languageEnglish (US)
Pages (from-to)105-114
Number of pages10
JournalJournal of Biomedical Materials Research
Volume38
Issue number2
DOIs
StatePublished - Jun 1997
Externally publishedYes

Keywords

  • biodegradable
  • pH
  • polyglycolic acid
  • polylactic acid
  • polymer

ASJC Scopus subject areas

  • Biomedical Engineering
  • Biomaterials

Fingerprint Dive into the research topics of 'Technique to control pH in vicinity of biodegrading PLA-PGA implants'. Together they form a unique fingerprint.

  • Cite this