SU‐FF‐T‐559

Effect of Cardiac Motion On the Cyberknife Synchrony Tracking System for Radiosurgical Cardiac Ablation

B. K. Teo, Sonja Dieterich, O. Blanck, T. Sumanaweera, E. Gardner

Research output: Contribution to journalArticle

3 Citations (Scopus)

Abstract

Purpose: To characterize the effect of cardiac motion on the CyberKnife Synchrony respiratory tracking system for radiosurgical cardiac ablation. Method and Materials: A high precision programmable haptic device (Force Dimensions Omega.6) was used to simulate cardiac and respiratory motion of a ball cube phantom. The CyberKnife Synchrony system was used to track the position of the 25 mm diameter target via spatial correlation of a fiducial within the phantom with the external LEDs placed on a second haptic device simulating chest wall motion. The targeting and tracking error were analyzed by comparing the actual target positions with the linac offset positions predicted from the Synchrony model. Film dosimetry was used to evaluate the effect of cardiac motion on target coverage. Results: Examination of log files revealed that the Synchrony modeling averaged out cardiac motion not present in the surrogate with some error. Superior‐inferior motion from both respiratory and cardiac motion contributed to the largest Synchrony modeling error. This modeling error led to misdirection of the beams from the time averaged target position. Additionally, untracked motion resulted in the target moving into and out of the radiation beams which manifested as softening of the dose gradient changes. Analysis of the films indicated that this was most severe in the superior inferior direction consistent with the extent of motion in that direction. Conclusion: A small additional margin would enable treatment of cardiac targets with the current CyberKnife if Synchrony tracking were used. For the motions and dose distributions tested, a 2 mm margin was sufficient to provide good target coverage in the presence of untracked cardiac motion. S. Dieterich consults for Cyberheart Inc.

Original languageEnglish (US)
Number of pages1
JournalMedical Physics
Volume36
Issue number6
DOIs
StatePublished - 2009
Externally publishedYes

Fingerprint

Film Dosimetry
Equipment and Supplies
Thoracic Wall
Respiratory System
Radiation
Direction compound

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

SU‐FF‐T‐559 : Effect of Cardiac Motion On the Cyberknife Synchrony Tracking System for Radiosurgical Cardiac Ablation. / Teo, B. K.; Dieterich, Sonja; Blanck, O.; Sumanaweera, T.; Gardner, E.

In: Medical Physics, Vol. 36, No. 6, 2009.

Research output: Contribution to journalArticle

@article{02b585610b014ab28406ff903f1dc401,
title = "SU‐FF‐T‐559: Effect of Cardiac Motion On the Cyberknife Synchrony Tracking System for Radiosurgical Cardiac Ablation",
abstract = "Purpose: To characterize the effect of cardiac motion on the CyberKnife Synchrony respiratory tracking system for radiosurgical cardiac ablation. Method and Materials: A high precision programmable haptic device (Force Dimensions Omega.6) was used to simulate cardiac and respiratory motion of a ball cube phantom. The CyberKnife Synchrony system was used to track the position of the 25 mm diameter target via spatial correlation of a fiducial within the phantom with the external LEDs placed on a second haptic device simulating chest wall motion. The targeting and tracking error were analyzed by comparing the actual target positions with the linac offset positions predicted from the Synchrony model. Film dosimetry was used to evaluate the effect of cardiac motion on target coverage. Results: Examination of log files revealed that the Synchrony modeling averaged out cardiac motion not present in the surrogate with some error. Superior‐inferior motion from both respiratory and cardiac motion contributed to the largest Synchrony modeling error. This modeling error led to misdirection of the beams from the time averaged target position. Additionally, untracked motion resulted in the target moving into and out of the radiation beams which manifested as softening of the dose gradient changes. Analysis of the films indicated that this was most severe in the superior inferior direction consistent with the extent of motion in that direction. Conclusion: A small additional margin would enable treatment of cardiac targets with the current CyberKnife if Synchrony tracking were used. For the motions and dose distributions tested, a 2 mm margin was sufficient to provide good target coverage in the presence of untracked cardiac motion. S. Dieterich consults for Cyberheart Inc.",
author = "Teo, {B. K.} and Sonja Dieterich and O. Blanck and T. Sumanaweera and E. Gardner",
year = "2009",
doi = "10.1118/1.3182057",
language = "English (US)",
volume = "36",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "6",

}

TY - JOUR

T1 - SU‐FF‐T‐559

T2 - Effect of Cardiac Motion On the Cyberknife Synchrony Tracking System for Radiosurgical Cardiac Ablation

AU - Teo, B. K.

AU - Dieterich, Sonja

AU - Blanck, O.

AU - Sumanaweera, T.

AU - Gardner, E.

PY - 2009

Y1 - 2009

N2 - Purpose: To characterize the effect of cardiac motion on the CyberKnife Synchrony respiratory tracking system for radiosurgical cardiac ablation. Method and Materials: A high precision programmable haptic device (Force Dimensions Omega.6) was used to simulate cardiac and respiratory motion of a ball cube phantom. The CyberKnife Synchrony system was used to track the position of the 25 mm diameter target via spatial correlation of a fiducial within the phantom with the external LEDs placed on a second haptic device simulating chest wall motion. The targeting and tracking error were analyzed by comparing the actual target positions with the linac offset positions predicted from the Synchrony model. Film dosimetry was used to evaluate the effect of cardiac motion on target coverage. Results: Examination of log files revealed that the Synchrony modeling averaged out cardiac motion not present in the surrogate with some error. Superior‐inferior motion from both respiratory and cardiac motion contributed to the largest Synchrony modeling error. This modeling error led to misdirection of the beams from the time averaged target position. Additionally, untracked motion resulted in the target moving into and out of the radiation beams which manifested as softening of the dose gradient changes. Analysis of the films indicated that this was most severe in the superior inferior direction consistent with the extent of motion in that direction. Conclusion: A small additional margin would enable treatment of cardiac targets with the current CyberKnife if Synchrony tracking were used. For the motions and dose distributions tested, a 2 mm margin was sufficient to provide good target coverage in the presence of untracked cardiac motion. S. Dieterich consults for Cyberheart Inc.

AB - Purpose: To characterize the effect of cardiac motion on the CyberKnife Synchrony respiratory tracking system for radiosurgical cardiac ablation. Method and Materials: A high precision programmable haptic device (Force Dimensions Omega.6) was used to simulate cardiac and respiratory motion of a ball cube phantom. The CyberKnife Synchrony system was used to track the position of the 25 mm diameter target via spatial correlation of a fiducial within the phantom with the external LEDs placed on a second haptic device simulating chest wall motion. The targeting and tracking error were analyzed by comparing the actual target positions with the linac offset positions predicted from the Synchrony model. Film dosimetry was used to evaluate the effect of cardiac motion on target coverage. Results: Examination of log files revealed that the Synchrony modeling averaged out cardiac motion not present in the surrogate with some error. Superior‐inferior motion from both respiratory and cardiac motion contributed to the largest Synchrony modeling error. This modeling error led to misdirection of the beams from the time averaged target position. Additionally, untracked motion resulted in the target moving into and out of the radiation beams which manifested as softening of the dose gradient changes. Analysis of the films indicated that this was most severe in the superior inferior direction consistent with the extent of motion in that direction. Conclusion: A small additional margin would enable treatment of cardiac targets with the current CyberKnife if Synchrony tracking were used. For the motions and dose distributions tested, a 2 mm margin was sufficient to provide good target coverage in the presence of untracked cardiac motion. S. Dieterich consults for Cyberheart Inc.

UR - http://www.scopus.com/inward/record.url?scp=84994633580&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84994633580&partnerID=8YFLogxK

U2 - 10.1118/1.3182057

DO - 10.1118/1.3182057

M3 - Article

VL - 36

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 6

ER -