Rate of change of oxygen concentration for a large animal circle anesthetic system

Adrian M. Solano, Robert J Brosnan, Eugene Steffey

Research output: Contribution to journalArticle

6 Citations (Scopus)

Abstract

Objective - To describe the effects of changes in circuit volume and oxygen inflow rate on inspired oxygen concentration for a large animal circle anesthetic system. Study population - A large animal circle anesthetic system, a 10 L/min flowmeter, and 20- and 40-L breathing bags. Procedure - Circuit volume was determined by a carbon dioxide dilution technique. Oxygen flow rates of 3, 6, and 10 L/min were delivered to the circuit with the large breathing bag, and a flow rate of 6 L/min was used with the small bag. Gas samples were collected during a 20-minute period. The time constant (τ) and half-time (T1/2) were calculated and compared with measured values. Results - Mean ± SEM volume of the breathing circuit with a 20- and 40-L breathing bag was 32.97 ± 0.91 L and 49.26 ± 0.58 L, respectively. The τ from measurements was 11.97, 6.10, and 3.60 minutes at oxygen flow rates of 3, 6, and 10 L/min, respectively, for the large breathing bag and 3.73 minutes at a flow rate of 6 L/min for the small breathing bag. The T1/2 was 8.29, 4.22, and 2.49 minutes at oxygen flow rates of 3, 6, and 10 L/min, respectively, for the large breathing bag and 2.58 minutes for the small breathing bag. Conclusions and clinical relevance - This study emphasizes that there are delays in the rate of increase in the inspired oxygen concentration that accompany use of conventional large animal circle anesthetic systems and low rates of inflow for fresh oxygen.

Original languageEnglish (US)
Pages (from-to)1675-1678
Number of pages4
JournalAmerican Journal of Veterinary Research
Volume66
Issue number10
DOIs
StatePublished - Oct 2005

Fingerprint

anesthetics
breathing
Anesthetics
bags
Respiration
Oxygen
oxygen
animals
Indicator Dilution Techniques
Flowmeters
flowmeters
Carbon Dioxide
Gases
carbon dioxide
gases
Population

ASJC Scopus subject areas

  • veterinary(all)

Cite this

Rate of change of oxygen concentration for a large animal circle anesthetic system. / Solano, Adrian M.; Brosnan, Robert J; Steffey, Eugene.

In: American Journal of Veterinary Research, Vol. 66, No. 10, 10.2005, p. 1675-1678.

Research output: Contribution to journalArticle

@article{b11465ce6281464395e2f3451def480a,
title = "Rate of change of oxygen concentration for a large animal circle anesthetic system",
abstract = "Objective - To describe the effects of changes in circuit volume and oxygen inflow rate on inspired oxygen concentration for a large animal circle anesthetic system. Study population - A large animal circle anesthetic system, a 10 L/min flowmeter, and 20- and 40-L breathing bags. Procedure - Circuit volume was determined by a carbon dioxide dilution technique. Oxygen flow rates of 3, 6, and 10 L/min were delivered to the circuit with the large breathing bag, and a flow rate of 6 L/min was used with the small bag. Gas samples were collected during a 20-minute period. The time constant (τ) and half-time (T1/2) were calculated and compared with measured values. Results - Mean ± SEM volume of the breathing circuit with a 20- and 40-L breathing bag was 32.97 ± 0.91 L and 49.26 ± 0.58 L, respectively. The τ from measurements was 11.97, 6.10, and 3.60 minutes at oxygen flow rates of 3, 6, and 10 L/min, respectively, for the large breathing bag and 3.73 minutes at a flow rate of 6 L/min for the small breathing bag. The T1/2 was 8.29, 4.22, and 2.49 minutes at oxygen flow rates of 3, 6, and 10 L/min, respectively, for the large breathing bag and 2.58 minutes for the small breathing bag. Conclusions and clinical relevance - This study emphasizes that there are delays in the rate of increase in the inspired oxygen concentration that accompany use of conventional large animal circle anesthetic systems and low rates of inflow for fresh oxygen.",
author = "Solano, {Adrian M.} and Brosnan, {Robert J} and Eugene Steffey",
year = "2005",
month = "10",
doi = "10.2460/ajvr.2005.66.1675",
language = "English (US)",
volume = "66",
pages = "1675--1678",
journal = "American Journal of Veterinary Research",
issn = "0002-9645",
publisher = "American Veterinary Medical Association",
number = "10",

}

TY - JOUR

T1 - Rate of change of oxygen concentration for a large animal circle anesthetic system

AU - Solano, Adrian M.

AU - Brosnan, Robert J

AU - Steffey, Eugene

PY - 2005/10

Y1 - 2005/10

N2 - Objective - To describe the effects of changes in circuit volume and oxygen inflow rate on inspired oxygen concentration for a large animal circle anesthetic system. Study population - A large animal circle anesthetic system, a 10 L/min flowmeter, and 20- and 40-L breathing bags. Procedure - Circuit volume was determined by a carbon dioxide dilution technique. Oxygen flow rates of 3, 6, and 10 L/min were delivered to the circuit with the large breathing bag, and a flow rate of 6 L/min was used with the small bag. Gas samples were collected during a 20-minute period. The time constant (τ) and half-time (T1/2) were calculated and compared with measured values. Results - Mean ± SEM volume of the breathing circuit with a 20- and 40-L breathing bag was 32.97 ± 0.91 L and 49.26 ± 0.58 L, respectively. The τ from measurements was 11.97, 6.10, and 3.60 minutes at oxygen flow rates of 3, 6, and 10 L/min, respectively, for the large breathing bag and 3.73 minutes at a flow rate of 6 L/min for the small breathing bag. The T1/2 was 8.29, 4.22, and 2.49 minutes at oxygen flow rates of 3, 6, and 10 L/min, respectively, for the large breathing bag and 2.58 minutes for the small breathing bag. Conclusions and clinical relevance - This study emphasizes that there are delays in the rate of increase in the inspired oxygen concentration that accompany use of conventional large animal circle anesthetic systems and low rates of inflow for fresh oxygen.

AB - Objective - To describe the effects of changes in circuit volume and oxygen inflow rate on inspired oxygen concentration for a large animal circle anesthetic system. Study population - A large animal circle anesthetic system, a 10 L/min flowmeter, and 20- and 40-L breathing bags. Procedure - Circuit volume was determined by a carbon dioxide dilution technique. Oxygen flow rates of 3, 6, and 10 L/min were delivered to the circuit with the large breathing bag, and a flow rate of 6 L/min was used with the small bag. Gas samples were collected during a 20-minute period. The time constant (τ) and half-time (T1/2) were calculated and compared with measured values. Results - Mean ± SEM volume of the breathing circuit with a 20- and 40-L breathing bag was 32.97 ± 0.91 L and 49.26 ± 0.58 L, respectively. The τ from measurements was 11.97, 6.10, and 3.60 minutes at oxygen flow rates of 3, 6, and 10 L/min, respectively, for the large breathing bag and 3.73 minutes at a flow rate of 6 L/min for the small breathing bag. The T1/2 was 8.29, 4.22, and 2.49 minutes at oxygen flow rates of 3, 6, and 10 L/min, respectively, for the large breathing bag and 2.58 minutes for the small breathing bag. Conclusions and clinical relevance - This study emphasizes that there are delays in the rate of increase in the inspired oxygen concentration that accompany use of conventional large animal circle anesthetic systems and low rates of inflow for fresh oxygen.

UR - http://www.scopus.com/inward/record.url?scp=27844602875&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=27844602875&partnerID=8YFLogxK

U2 - 10.2460/ajvr.2005.66.1675

DO - 10.2460/ajvr.2005.66.1675

M3 - Article

VL - 66

SP - 1675

EP - 1678

JO - American Journal of Veterinary Research

JF - American Journal of Veterinary Research

SN - 0002-9645

IS - 10

ER -