Radiographic evaluation of acetabular component position in dogs

Jonathan Dyce, Erik R Wisner, Steven C. Schrader, Qiang Wang, Marvin L. Olmstead

Research output: Contribution to journalArticle

30 Citations (Scopus)

Abstract

Objective - To assess the usefulness of standard radiographic projections for determining acetabular cup position and to compare 2 radiographic methods for calculation of the angle of lateral opening of the acetabular cup. Study Design - In vitro radiographic study. Methods - Acetabular components were mounted on a custom-built inclinometer. The effect of varying the angles of lateral opening, inclination, version, and centering of the x-ray beam on the radiographic appearance of the cup was investigated. The angle of lateral opening was calculated trigonometrically after direct measurement and by means of computer-aided image analysis. Results - Using a ventrodorsal radiograph, the computer-aided measurement of angle of lateral opening was accurate to within 5° when the true angle of lateral opening was ≥40°. In the case of direct measurement, the calculated angle was accurate to within 5° only when the true angle was ≥60°. An increasing angle of inclination was associated with reduced accuracy of calculation of the angles of lateral opening and version. Standard radiographic assessment of acetabular version and inclination was not found to be clinically useful. The displacement of the x-ray beam, which would cause significant error in the determination of lateral opening, is unlikely to be exceeded in practice. Cup size had no significant effect on the accuracy of the results. Conclusions - Computer-aided image analysis can be used to calculate angles of lateral opening to within 10° throughout the range of acetabular positions seen in practice and is appropriate for postoperative documentation of cup position. The angles of inclination and version should be interpreted with caution.

Original languageEnglish (US)
Pages (from-to)28-39
Number of pages12
JournalVeterinary Surgery
Volume30
Issue number1
DOIs
StatePublished - Jan 2001
Externally publishedYes

Fingerprint

Dogs
dogs
X-radiation
X-Rays
image analysis
Documentation
in vitro studies
experimental design
methodology
In Vitro Techniques

ASJC Scopus subject areas

  • veterinary(all)

Cite this

Radiographic evaluation of acetabular component position in dogs. / Dyce, Jonathan; Wisner, Erik R; Schrader, Steven C.; Wang, Qiang; Olmstead, Marvin L.

In: Veterinary Surgery, Vol. 30, No. 1, 01.2001, p. 28-39.

Research output: Contribution to journalArticle

Dyce, Jonathan ; Wisner, Erik R ; Schrader, Steven C. ; Wang, Qiang ; Olmstead, Marvin L. / Radiographic evaluation of acetabular component position in dogs. In: Veterinary Surgery. 2001 ; Vol. 30, No. 1. pp. 28-39.
@article{44a9c0fd8c6f43379255ecfa3b98a367,
title = "Radiographic evaluation of acetabular component position in dogs",
abstract = "Objective - To assess the usefulness of standard radiographic projections for determining acetabular cup position and to compare 2 radiographic methods for calculation of the angle of lateral opening of the acetabular cup. Study Design - In vitro radiographic study. Methods - Acetabular components were mounted on a custom-built inclinometer. The effect of varying the angles of lateral opening, inclination, version, and centering of the x-ray beam on the radiographic appearance of the cup was investigated. The angle of lateral opening was calculated trigonometrically after direct measurement and by means of computer-aided image analysis. Results - Using a ventrodorsal radiograph, the computer-aided measurement of angle of lateral opening was accurate to within 5° when the true angle of lateral opening was ≥40°. In the case of direct measurement, the calculated angle was accurate to within 5° only when the true angle was ≥60°. An increasing angle of inclination was associated with reduced accuracy of calculation of the angles of lateral opening and version. Standard radiographic assessment of acetabular version and inclination was not found to be clinically useful. The displacement of the x-ray beam, which would cause significant error in the determination of lateral opening, is unlikely to be exceeded in practice. Cup size had no significant effect on the accuracy of the results. Conclusions - Computer-aided image analysis can be used to calculate angles of lateral opening to within 10° throughout the range of acetabular positions seen in practice and is appropriate for postoperative documentation of cup position. The angles of inclination and version should be interpreted with caution.",
author = "Jonathan Dyce and Wisner, {Erik R} and Schrader, {Steven C.} and Qiang Wang and Olmstead, {Marvin L.}",
year = "2001",
month = "1",
doi = "10.1053/jvet.2001.20343",
language = "English (US)",
volume = "30",
pages = "28--39",
journal = "Veterinary Surgery",
issn = "0161-3499",
publisher = "Wiley-Blackwell",
number = "1",

}

TY - JOUR

T1 - Radiographic evaluation of acetabular component position in dogs

AU - Dyce, Jonathan

AU - Wisner, Erik R

AU - Schrader, Steven C.

AU - Wang, Qiang

AU - Olmstead, Marvin L.

PY - 2001/1

Y1 - 2001/1

N2 - Objective - To assess the usefulness of standard radiographic projections for determining acetabular cup position and to compare 2 radiographic methods for calculation of the angle of lateral opening of the acetabular cup. Study Design - In vitro radiographic study. Methods - Acetabular components were mounted on a custom-built inclinometer. The effect of varying the angles of lateral opening, inclination, version, and centering of the x-ray beam on the radiographic appearance of the cup was investigated. The angle of lateral opening was calculated trigonometrically after direct measurement and by means of computer-aided image analysis. Results - Using a ventrodorsal radiograph, the computer-aided measurement of angle of lateral opening was accurate to within 5° when the true angle of lateral opening was ≥40°. In the case of direct measurement, the calculated angle was accurate to within 5° only when the true angle was ≥60°. An increasing angle of inclination was associated with reduced accuracy of calculation of the angles of lateral opening and version. Standard radiographic assessment of acetabular version and inclination was not found to be clinically useful. The displacement of the x-ray beam, which would cause significant error in the determination of lateral opening, is unlikely to be exceeded in practice. Cup size had no significant effect on the accuracy of the results. Conclusions - Computer-aided image analysis can be used to calculate angles of lateral opening to within 10° throughout the range of acetabular positions seen in practice and is appropriate for postoperative documentation of cup position. The angles of inclination and version should be interpreted with caution.

AB - Objective - To assess the usefulness of standard radiographic projections for determining acetabular cup position and to compare 2 radiographic methods for calculation of the angle of lateral opening of the acetabular cup. Study Design - In vitro radiographic study. Methods - Acetabular components were mounted on a custom-built inclinometer. The effect of varying the angles of lateral opening, inclination, version, and centering of the x-ray beam on the radiographic appearance of the cup was investigated. The angle of lateral opening was calculated trigonometrically after direct measurement and by means of computer-aided image analysis. Results - Using a ventrodorsal radiograph, the computer-aided measurement of angle of lateral opening was accurate to within 5° when the true angle of lateral opening was ≥40°. In the case of direct measurement, the calculated angle was accurate to within 5° only when the true angle was ≥60°. An increasing angle of inclination was associated with reduced accuracy of calculation of the angles of lateral opening and version. Standard radiographic assessment of acetabular version and inclination was not found to be clinically useful. The displacement of the x-ray beam, which would cause significant error in the determination of lateral opening, is unlikely to be exceeded in practice. Cup size had no significant effect on the accuracy of the results. Conclusions - Computer-aided image analysis can be used to calculate angles of lateral opening to within 10° throughout the range of acetabular positions seen in practice and is appropriate for postoperative documentation of cup position. The angles of inclination and version should be interpreted with caution.

UR - http://www.scopus.com/inward/record.url?scp=0035220136&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0035220136&partnerID=8YFLogxK

U2 - 10.1053/jvet.2001.20343

DO - 10.1053/jvet.2001.20343

M3 - Article

C2 - 11172458

AN - SCOPUS:0035220136

VL - 30

SP - 28

EP - 39

JO - Veterinary Surgery

JF - Veterinary Surgery

SN - 0161-3499

IS - 1

ER -