Quantitative, simultaneous PET/MRI for intratumoral imaging with an MRI-compatible PET scanner

Thomas S C Ng, James R. Bading, Ryan Park, Hargun Sohi, Daniel Procissi, David Colcher, Peter S. Conti, Simon R Cherry, Andrew A. Raubitschek, Russell E. Jacobs

Research output: Contribution to journalArticle

20 Citations (Scopus)

Abstract

Noninvasive methods are needed to explore the heterogeneous tumor microenvironment and its modulation by therapy. Hybrid PET/MRI systems are being developed for small-animal and clinical use. The advantage of these integrated systems depends on their ability to provide MR images that are spatially coincident with simultaneously acquired PET images, allowing combined functional MRI and PET studies of intratissue heterogeneity. Although much effort has been devoted to developing this new technology, the issue of quantitative and spatial fidelity of PET images from hybrid PET/MRI systems to the tissues imaged has received little attention. Here, we evaluated the ability of a first-generation, small-animal MRI-compatible PET scanner to accurately depict heterogeneous patterns of radiotracer uptake in tumors. Methods: Quantitative imaging characteristics of the MRI-compatible PET (PET/MRI) scanner were evaluated with phantoms using calibration coefficients derived from a mouse-sized linearity phantom. PET performance was compared with a commercial small-animal PET system and autoradiography in tumor-bearing mice. Pixel and structure-based similarity metrics were used to evaluate image concordance among modalities. Feasibility of simultaneous PET/MRI functional imaging of tumors was explored by following 64Cu-labeled antibody uptake in relation to diffusion MRI using cooccurrence matrix analysis. Results: The PET/MRI scanner showed stable and linear response. Activity concentration recovery values (measured and true activity concentration) calculated for 4-mm-diameter rods within linearity and uniform activity rod phantoms were near unity (0.97 ± 0.06 and 1.03 ± 0.03, respectively). Intratumoral uptake patterns for both 18F-FDG and a 64Cu-antibody acquired using the PET/MRI scanner and small-animal PET were highly correlated with autoradiography (r > 0.99) and with each other (r = 0.97 ± 0.01). On the basis of these data, we performed a preliminary study comparing diffusion MRI and radiolabeled antibody uptake patterns over time and visualized movement of antibodies from the vascular space into the tumor mass. Conclusion: The MRI-compatible PET scanner provided tumor images that were quantitatively accurate and spatially concordant with autoradiography and the small-animal PET examination. Cooccurrence matrix approaches enabled effective analysis of multimodal image sets. These observations confirm the ability of the current simultaneous PET/MRI system to provide accurate observations of intratumoral function and serve as a benchmark for future evaluations of hybrid instrumentation.

Original languageEnglish (US)
Pages (from-to)1102-1109
Number of pages8
JournalJournal of Nuclear Medicine
Volume53
Issue number7
DOIs
StatePublished - Jul 1 2012

Fingerprint

Autoradiography
Diffusion Magnetic Resonance Imaging
Antibodies
Neoplasms
Magnetic Resonance Imaging
Benchmarking
Tumor Microenvironment
Fluorodeoxyglucose F18
Calibration
Blood Vessels
Technology
Therapeutics

Keywords

  • Multimodal imaging
  • PET/MRI
  • Preclinical
  • Quantitative molecular imaging
  • Tumor heterogeneity

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Cite this

Ng, T. S. C., Bading, J. R., Park, R., Sohi, H., Procissi, D., Colcher, D., ... Jacobs, R. E. (2012). Quantitative, simultaneous PET/MRI for intratumoral imaging with an MRI-compatible PET scanner. Journal of Nuclear Medicine, 53(7), 1102-1109. https://doi.org/10.2967/jnumed.111.099861

Quantitative, simultaneous PET/MRI for intratumoral imaging with an MRI-compatible PET scanner. / Ng, Thomas S C; Bading, James R.; Park, Ryan; Sohi, Hargun; Procissi, Daniel; Colcher, David; Conti, Peter S.; Cherry, Simon R; Raubitschek, Andrew A.; Jacobs, Russell E.

In: Journal of Nuclear Medicine, Vol. 53, No. 7, 01.07.2012, p. 1102-1109.

Research output: Contribution to journalArticle

Ng, TSC, Bading, JR, Park, R, Sohi, H, Procissi, D, Colcher, D, Conti, PS, Cherry, SR, Raubitschek, AA & Jacobs, RE 2012, 'Quantitative, simultaneous PET/MRI for intratumoral imaging with an MRI-compatible PET scanner', Journal of Nuclear Medicine, vol. 53, no. 7, pp. 1102-1109. https://doi.org/10.2967/jnumed.111.099861
Ng, Thomas S C ; Bading, James R. ; Park, Ryan ; Sohi, Hargun ; Procissi, Daniel ; Colcher, David ; Conti, Peter S. ; Cherry, Simon R ; Raubitschek, Andrew A. ; Jacobs, Russell E. / Quantitative, simultaneous PET/MRI for intratumoral imaging with an MRI-compatible PET scanner. In: Journal of Nuclear Medicine. 2012 ; Vol. 53, No. 7. pp. 1102-1109.
@article{3d6f055c7b2e4dee919d100694a9f52c,
title = "Quantitative, simultaneous PET/MRI for intratumoral imaging with an MRI-compatible PET scanner",
abstract = "Noninvasive methods are needed to explore the heterogeneous tumor microenvironment and its modulation by therapy. Hybrid PET/MRI systems are being developed for small-animal and clinical use. The advantage of these integrated systems depends on their ability to provide MR images that are spatially coincident with simultaneously acquired PET images, allowing combined functional MRI and PET studies of intratissue heterogeneity. Although much effort has been devoted to developing this new technology, the issue of quantitative and spatial fidelity of PET images from hybrid PET/MRI systems to the tissues imaged has received little attention. Here, we evaluated the ability of a first-generation, small-animal MRI-compatible PET scanner to accurately depict heterogeneous patterns of radiotracer uptake in tumors. Methods: Quantitative imaging characteristics of the MRI-compatible PET (PET/MRI) scanner were evaluated with phantoms using calibration coefficients derived from a mouse-sized linearity phantom. PET performance was compared with a commercial small-animal PET system and autoradiography in tumor-bearing mice. Pixel and structure-based similarity metrics were used to evaluate image concordance among modalities. Feasibility of simultaneous PET/MRI functional imaging of tumors was explored by following 64Cu-labeled antibody uptake in relation to diffusion MRI using cooccurrence matrix analysis. Results: The PET/MRI scanner showed stable and linear response. Activity concentration recovery values (measured and true activity concentration) calculated for 4-mm-diameter rods within linearity and uniform activity rod phantoms were near unity (0.97 ± 0.06 and 1.03 ± 0.03, respectively). Intratumoral uptake patterns for both 18F-FDG and a 64Cu-antibody acquired using the PET/MRI scanner and small-animal PET were highly correlated with autoradiography (r > 0.99) and with each other (r = 0.97 ± 0.01). On the basis of these data, we performed a preliminary study comparing diffusion MRI and radiolabeled antibody uptake patterns over time and visualized movement of antibodies from the vascular space into the tumor mass. Conclusion: The MRI-compatible PET scanner provided tumor images that were quantitatively accurate and spatially concordant with autoradiography and the small-animal PET examination. Cooccurrence matrix approaches enabled effective analysis of multimodal image sets. These observations confirm the ability of the current simultaneous PET/MRI system to provide accurate observations of intratumoral function and serve as a benchmark for future evaluations of hybrid instrumentation.",
keywords = "Multimodal imaging, PET/MRI, Preclinical, Quantitative molecular imaging, Tumor heterogeneity",
author = "Ng, {Thomas S C} and Bading, {James R.} and Ryan Park and Hargun Sohi and Daniel Procissi and David Colcher and Conti, {Peter S.} and Cherry, {Simon R} and Raubitschek, {Andrew A.} and Jacobs, {Russell E.}",
year = "2012",
month = "7",
day = "1",
doi = "10.2967/jnumed.111.099861",
language = "English (US)",
volume = "53",
pages = "1102--1109",
journal = "Journal of Nuclear Medicine",
issn = "0161-5505",
publisher = "Society of Nuclear Medicine Inc.",
number = "7",

}

TY - JOUR

T1 - Quantitative, simultaneous PET/MRI for intratumoral imaging with an MRI-compatible PET scanner

AU - Ng, Thomas S C

AU - Bading, James R.

AU - Park, Ryan

AU - Sohi, Hargun

AU - Procissi, Daniel

AU - Colcher, David

AU - Conti, Peter S.

AU - Cherry, Simon R

AU - Raubitschek, Andrew A.

AU - Jacobs, Russell E.

PY - 2012/7/1

Y1 - 2012/7/1

N2 - Noninvasive methods are needed to explore the heterogeneous tumor microenvironment and its modulation by therapy. Hybrid PET/MRI systems are being developed for small-animal and clinical use. The advantage of these integrated systems depends on their ability to provide MR images that are spatially coincident with simultaneously acquired PET images, allowing combined functional MRI and PET studies of intratissue heterogeneity. Although much effort has been devoted to developing this new technology, the issue of quantitative and spatial fidelity of PET images from hybrid PET/MRI systems to the tissues imaged has received little attention. Here, we evaluated the ability of a first-generation, small-animal MRI-compatible PET scanner to accurately depict heterogeneous patterns of radiotracer uptake in tumors. Methods: Quantitative imaging characteristics of the MRI-compatible PET (PET/MRI) scanner were evaluated with phantoms using calibration coefficients derived from a mouse-sized linearity phantom. PET performance was compared with a commercial small-animal PET system and autoradiography in tumor-bearing mice. Pixel and structure-based similarity metrics were used to evaluate image concordance among modalities. Feasibility of simultaneous PET/MRI functional imaging of tumors was explored by following 64Cu-labeled antibody uptake in relation to diffusion MRI using cooccurrence matrix analysis. Results: The PET/MRI scanner showed stable and linear response. Activity concentration recovery values (measured and true activity concentration) calculated for 4-mm-diameter rods within linearity and uniform activity rod phantoms were near unity (0.97 ± 0.06 and 1.03 ± 0.03, respectively). Intratumoral uptake patterns for both 18F-FDG and a 64Cu-antibody acquired using the PET/MRI scanner and small-animal PET were highly correlated with autoradiography (r > 0.99) and with each other (r = 0.97 ± 0.01). On the basis of these data, we performed a preliminary study comparing diffusion MRI and radiolabeled antibody uptake patterns over time and visualized movement of antibodies from the vascular space into the tumor mass. Conclusion: The MRI-compatible PET scanner provided tumor images that were quantitatively accurate and spatially concordant with autoradiography and the small-animal PET examination. Cooccurrence matrix approaches enabled effective analysis of multimodal image sets. These observations confirm the ability of the current simultaneous PET/MRI system to provide accurate observations of intratumoral function and serve as a benchmark for future evaluations of hybrid instrumentation.

AB - Noninvasive methods are needed to explore the heterogeneous tumor microenvironment and its modulation by therapy. Hybrid PET/MRI systems are being developed for small-animal and clinical use. The advantage of these integrated systems depends on their ability to provide MR images that are spatially coincident with simultaneously acquired PET images, allowing combined functional MRI and PET studies of intratissue heterogeneity. Although much effort has been devoted to developing this new technology, the issue of quantitative and spatial fidelity of PET images from hybrid PET/MRI systems to the tissues imaged has received little attention. Here, we evaluated the ability of a first-generation, small-animal MRI-compatible PET scanner to accurately depict heterogeneous patterns of radiotracer uptake in tumors. Methods: Quantitative imaging characteristics of the MRI-compatible PET (PET/MRI) scanner were evaluated with phantoms using calibration coefficients derived from a mouse-sized linearity phantom. PET performance was compared with a commercial small-animal PET system and autoradiography in tumor-bearing mice. Pixel and structure-based similarity metrics were used to evaluate image concordance among modalities. Feasibility of simultaneous PET/MRI functional imaging of tumors was explored by following 64Cu-labeled antibody uptake in relation to diffusion MRI using cooccurrence matrix analysis. Results: The PET/MRI scanner showed stable and linear response. Activity concentration recovery values (measured and true activity concentration) calculated for 4-mm-diameter rods within linearity and uniform activity rod phantoms were near unity (0.97 ± 0.06 and 1.03 ± 0.03, respectively). Intratumoral uptake patterns for both 18F-FDG and a 64Cu-antibody acquired using the PET/MRI scanner and small-animal PET were highly correlated with autoradiography (r > 0.99) and with each other (r = 0.97 ± 0.01). On the basis of these data, we performed a preliminary study comparing diffusion MRI and radiolabeled antibody uptake patterns over time and visualized movement of antibodies from the vascular space into the tumor mass. Conclusion: The MRI-compatible PET scanner provided tumor images that were quantitatively accurate and spatially concordant with autoradiography and the small-animal PET examination. Cooccurrence matrix approaches enabled effective analysis of multimodal image sets. These observations confirm the ability of the current simultaneous PET/MRI system to provide accurate observations of intratumoral function and serve as a benchmark for future evaluations of hybrid instrumentation.

KW - Multimodal imaging

KW - PET/MRI

KW - Preclinical

KW - Quantitative molecular imaging

KW - Tumor heterogeneity

UR - http://www.scopus.com/inward/record.url?scp=84863441382&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84863441382&partnerID=8YFLogxK

U2 - 10.2967/jnumed.111.099861

DO - 10.2967/jnumed.111.099861

M3 - Article

C2 - 22661534

AN - SCOPUS:84863441382

VL - 53

SP - 1102

EP - 1109

JO - Journal of Nuclear Medicine

JF - Journal of Nuclear Medicine

SN - 0161-5505

IS - 7

ER -