Optimization and performance evaluation of the microPET II scanner for in vivo small-animal imaging

Yongfeng Yang, Yuan Chuan Tai, Stefan Siegel, Danny F. Newport, Bing Bai, Quanzheng Li, Richard M. Leahy, Simon R Cherry

Research output: Contribution to journalArticle

125 Citations (Scopus)

Abstract

MicroPET II is a newly developed PET (positron emission tomography) scanner designed for high-resolution imaging of small animals. It consist of 17 640 LSO crystals each measuring 0.975 × 0.975 × 12.5 mm3, which are arranged in 42 contiguous rings, with 420 crystals per ring. The scanner has an axial field of view (FOV) of 4.9 cm and a transaxial FOV of 8.5 cm. The purpose of this study was to carefully evaluate the performance of the system and to optimize settings for in vivo mouse and rat imaging studies. The volumetric image resolution was found to depend strongly on the reconstruction algorithm employed and averaged 1.1 mm (1.4 μl) across the central 3 cm of the transaxial FOV when using a statistical reconstruction algorithm with accurate system modelling. The sensitivity, scatter fraction and noise-equivalent count (NEC) rate for mouse- and rat-sized phantoms were measured for different energy and timing windows. Mouse imaging was optimized with a wide open energy window (150-750 keV) and a 10 ns timing window, leading to a sensitivity of 3.3% at the centre of the FOV and a peak NEC rate of 235 000 cps for a total activity of 80 MBq (2.2 mCi) in the phantom. Rat imaging, due to the higher scatter fraction, and the activity that lies outside of the field of view, achieved a maximum NEC rate of 24 600 cps for a total activity of 80 MBq (2.2 mCi) in the phantom, with an energy window of 250-750 keV and a 6 ns timing window. The sensitivity at the centre of the FOV for these settings is 2.1%. This work demonstrates that different scanner settings are necessary to optimize the NEC count rate for different-sized animals and different injected doses. Finally, phantom and in vivo animal studies are presented to demonstrate the capabilities of microPET II for small-animal imaging studies.

Original languageEnglish (US)
Pages (from-to)2527-2545
Number of pages19
JournalPhysics in Medicine and Biology
Volume49
Issue number12
DOIs
StatePublished - Jun 21 2004

Fingerprint

scanners
field of view
animals
Noise
Animals
Imaging techniques
optimization
evaluation
Rats
rats
mice
time measurement
sensitivity
Crystals
Positron emission tomography
Positron-Emission Tomography
Image resolution
rings
image resolution
crystals

ASJC Scopus subject areas

  • Biomedical Engineering
  • Physics and Astronomy (miscellaneous)
  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

Optimization and performance evaluation of the microPET II scanner for in vivo small-animal imaging. / Yang, Yongfeng; Tai, Yuan Chuan; Siegel, Stefan; Newport, Danny F.; Bai, Bing; Li, Quanzheng; Leahy, Richard M.; Cherry, Simon R.

In: Physics in Medicine and Biology, Vol. 49, No. 12, 21.06.2004, p. 2527-2545.

Research output: Contribution to journalArticle

Yang, Yongfeng ; Tai, Yuan Chuan ; Siegel, Stefan ; Newport, Danny F. ; Bai, Bing ; Li, Quanzheng ; Leahy, Richard M. ; Cherry, Simon R. / Optimization and performance evaluation of the microPET II scanner for in vivo small-animal imaging. In: Physics in Medicine and Biology. 2004 ; Vol. 49, No. 12. pp. 2527-2545.
@article{18446d323c90475cabfbbf2d698abe22,
title = "Optimization and performance evaluation of the microPET II scanner for in vivo small-animal imaging",
abstract = "MicroPET II is a newly developed PET (positron emission tomography) scanner designed for high-resolution imaging of small animals. It consist of 17 640 LSO crystals each measuring 0.975 × 0.975 × 12.5 mm3, which are arranged in 42 contiguous rings, with 420 crystals per ring. The scanner has an axial field of view (FOV) of 4.9 cm and a transaxial FOV of 8.5 cm. The purpose of this study was to carefully evaluate the performance of the system and to optimize settings for in vivo mouse and rat imaging studies. The volumetric image resolution was found to depend strongly on the reconstruction algorithm employed and averaged 1.1 mm (1.4 μl) across the central 3 cm of the transaxial FOV when using a statistical reconstruction algorithm with accurate system modelling. The sensitivity, scatter fraction and noise-equivalent count (NEC) rate for mouse- and rat-sized phantoms were measured for different energy and timing windows. Mouse imaging was optimized with a wide open energy window (150-750 keV) and a 10 ns timing window, leading to a sensitivity of 3.3{\%} at the centre of the FOV and a peak NEC rate of 235 000 cps for a total activity of 80 MBq (2.2 mCi) in the phantom. Rat imaging, due to the higher scatter fraction, and the activity that lies outside of the field of view, achieved a maximum NEC rate of 24 600 cps for a total activity of 80 MBq (2.2 mCi) in the phantom, with an energy window of 250-750 keV and a 6 ns timing window. The sensitivity at the centre of the FOV for these settings is 2.1{\%}. This work demonstrates that different scanner settings are necessary to optimize the NEC count rate for different-sized animals and different injected doses. Finally, phantom and in vivo animal studies are presented to demonstrate the capabilities of microPET II for small-animal imaging studies.",
author = "Yongfeng Yang and Tai, {Yuan Chuan} and Stefan Siegel and Newport, {Danny F.} and Bing Bai and Quanzheng Li and Leahy, {Richard M.} and Cherry, {Simon R}",
year = "2004",
month = "6",
day = "21",
doi = "10.1088/0031-9155/49/12/005",
language = "English (US)",
volume = "49",
pages = "2527--2545",
journal = "Physics in Medicine and Biology",
issn = "0031-9155",
publisher = "IOP Publishing Ltd.",
number = "12",

}

TY - JOUR

T1 - Optimization and performance evaluation of the microPET II scanner for in vivo small-animal imaging

AU - Yang, Yongfeng

AU - Tai, Yuan Chuan

AU - Siegel, Stefan

AU - Newport, Danny F.

AU - Bai, Bing

AU - Li, Quanzheng

AU - Leahy, Richard M.

AU - Cherry, Simon R

PY - 2004/6/21

Y1 - 2004/6/21

N2 - MicroPET II is a newly developed PET (positron emission tomography) scanner designed for high-resolution imaging of small animals. It consist of 17 640 LSO crystals each measuring 0.975 × 0.975 × 12.5 mm3, which are arranged in 42 contiguous rings, with 420 crystals per ring. The scanner has an axial field of view (FOV) of 4.9 cm and a transaxial FOV of 8.5 cm. The purpose of this study was to carefully evaluate the performance of the system and to optimize settings for in vivo mouse and rat imaging studies. The volumetric image resolution was found to depend strongly on the reconstruction algorithm employed and averaged 1.1 mm (1.4 μl) across the central 3 cm of the transaxial FOV when using a statistical reconstruction algorithm with accurate system modelling. The sensitivity, scatter fraction and noise-equivalent count (NEC) rate for mouse- and rat-sized phantoms were measured for different energy and timing windows. Mouse imaging was optimized with a wide open energy window (150-750 keV) and a 10 ns timing window, leading to a sensitivity of 3.3% at the centre of the FOV and a peak NEC rate of 235 000 cps for a total activity of 80 MBq (2.2 mCi) in the phantom. Rat imaging, due to the higher scatter fraction, and the activity that lies outside of the field of view, achieved a maximum NEC rate of 24 600 cps for a total activity of 80 MBq (2.2 mCi) in the phantom, with an energy window of 250-750 keV and a 6 ns timing window. The sensitivity at the centre of the FOV for these settings is 2.1%. This work demonstrates that different scanner settings are necessary to optimize the NEC count rate for different-sized animals and different injected doses. Finally, phantom and in vivo animal studies are presented to demonstrate the capabilities of microPET II for small-animal imaging studies.

AB - MicroPET II is a newly developed PET (positron emission tomography) scanner designed for high-resolution imaging of small animals. It consist of 17 640 LSO crystals each measuring 0.975 × 0.975 × 12.5 mm3, which are arranged in 42 contiguous rings, with 420 crystals per ring. The scanner has an axial field of view (FOV) of 4.9 cm and a transaxial FOV of 8.5 cm. The purpose of this study was to carefully evaluate the performance of the system and to optimize settings for in vivo mouse and rat imaging studies. The volumetric image resolution was found to depend strongly on the reconstruction algorithm employed and averaged 1.1 mm (1.4 μl) across the central 3 cm of the transaxial FOV when using a statistical reconstruction algorithm with accurate system modelling. The sensitivity, scatter fraction and noise-equivalent count (NEC) rate for mouse- and rat-sized phantoms were measured for different energy and timing windows. Mouse imaging was optimized with a wide open energy window (150-750 keV) and a 10 ns timing window, leading to a sensitivity of 3.3% at the centre of the FOV and a peak NEC rate of 235 000 cps for a total activity of 80 MBq (2.2 mCi) in the phantom. Rat imaging, due to the higher scatter fraction, and the activity that lies outside of the field of view, achieved a maximum NEC rate of 24 600 cps for a total activity of 80 MBq (2.2 mCi) in the phantom, with an energy window of 250-750 keV and a 6 ns timing window. The sensitivity at the centre of the FOV for these settings is 2.1%. This work demonstrates that different scanner settings are necessary to optimize the NEC count rate for different-sized animals and different injected doses. Finally, phantom and in vivo animal studies are presented to demonstrate the capabilities of microPET II for small-animal imaging studies.

UR - http://www.scopus.com/inward/record.url?scp=3242789549&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=3242789549&partnerID=8YFLogxK

U2 - 10.1088/0031-9155/49/12/005

DO - 10.1088/0031-9155/49/12/005

M3 - Article

VL - 49

SP - 2527

EP - 2545

JO - Physics in Medicine and Biology

JF - Physics in Medicine and Biology

SN - 0031-9155

IS - 12

ER -