Needle aspiration lung biopsy: Reevaluation of the blood patch technique in an equine model

Elizabeth H Moore, D. K. Shelton, Erik R Wisner, M. L. Richardson, D. M. Bishop, J. M. Brock

Research output: Contribution to journalArticle

17 Scopus citations

Abstract

PURPOSE: To reexamine the blood patch technique in a laboratory model of lung biopsy free of confounding clinical variables. MATERIALS AND METHODS: An equine model of lung biopsy was developed with an excised lobe connected to an insufflation bulb and pressure monitor. Patched and control unpatched punctures were made in the lung surface, and the pressure within the lung was raised to the maximum achievable. Whether air leakage from the puncture sites could be induced was determined and, if so, at what pressure it occurred. RESULTS: At statistical analysis with the Kaplan-Meier test and the Cox proportional hazards regression model, the difference between failure of the patched and unpatched punctures was statistically significant (P < .0001). CONCLUSION: The blood patch technique is effective in the laboratory setting and deserves reevaluation in a clinical series with updated biopsy techniques.

Original languageEnglish (US)
Pages (from-to)183-186
Number of pages4
JournalRadiology
Volume196
Issue number1
StatePublished - 1995

Keywords

  • Biopsies, complications
  • Biopsies, technology
  • Lung, biopsy
  • Pneumothorax

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology

Fingerprint Dive into the research topics of 'Needle aspiration lung biopsy: Reevaluation of the blood patch technique in an equine model'. Together they form a unique fingerprint.

  • Cite this

    Moore, E. H., Shelton, D. K., Wisner, E. R., Richardson, M. L., Bishop, D. M., & Brock, J. M. (1995). Needle aspiration lung biopsy: Reevaluation of the blood patch technique in an equine model. Radiology, 196(1), 183-186.