Long-term safety and durability of percutaneous septal sinus shortening (The PS3 system™) in an ovine model

Jason H Rogers, David A. Rahdert, Gary R Caputo, Patricia A. Takeda, Igor F. Palacios, Fermin O. Tio, Elizabeth A. Taylor, Reginald Low

Research output: Contribution to journalArticlepeer-review

12 Scopus citations

Abstract

Objectives: Chronic implants of the PS3TM system were conducted in an ovine model to assess durability and safety at up to 1 year follow-up. Background: The long-term durability and safety of emerging percutaneous devices for functional mitral regurgitation remain largely unknown. Methods: The PS3 system (consisting of interatrial septal and great cardiac vein devices connected by an adjustable suture bridge) was placed in eight healthy adult sheep. The mitral annular septal-lateral dimension in systole (SLS) was acutely reduced by 15-20%. Animals were sacrificed at up to 12 months postimplant and characterized by intracardiac echocardiography, cardiac computed tomography (CT), and histopathology. In vivo forces exerted on the PS3 bridge were measured by means of a novel load cell catheter. Results: At 3, 6, and 12 months after implantation, intracardiac echocardiographic and CT showed the PS3 systems to be intact without erosion and with overall sustained reductions in the SLS. Histopathologic assessment revealed each component correctly deployed in its respective target site without evidence of erosion, thrombus, or device fracture. The SLS was 26.5 ± 1.7 mm preimplant, 22.0 ± 1.4 mm post-PS3 (17.0% reduction), and 22.0 6 2.1 mm at latest follow-up. Mean forces exerted on the bridge in vivo ranged from 1.16 N to 1.87 N. Conclusions: The PS3 System demonstrated excellent biocompatibility without evidence of erosion, thrombosis, or perforation at up to one-year follow-up in this chronic healthy ovine model. Forces exerted in the PS3 system were relatively modest and should contribute to the durability of the device.

Original languageEnglish (US)
Pages (from-to)540-548
Number of pages9
JournalCatheterization and Cardiovascular Interventions
Volume73
Issue number4
DOIs
StatePublished - Mar 1 2009

Keywords

  • Intracardiac echo
  • Transeptal cath
  • Valvular heart disease

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Long-term safety and durability of percutaneous septal sinus shortening (The PS<sup>3</sup> system™) in an ovine model'. Together they form a unique fingerprint.

Cite this