LC-MS/MS screen for penitrem A and roquefortine C in serum and urine samples

Elizabeth R. Tor, Birgit Puschner, Michael S. Filigenzi, Asheesh K. Tiwary, Robert H Poppenga

Research output: Contribution to journalArticle

12 Citations (Scopus)

Abstract

A rapid LC-MS/MS method, using a triple quadrupole/linear ion trap mass spectrometer, was developed for determination of penitrem A and roquefortine C in serum and urine samples. Penitrem A and roquefortine C were extracted from samples with methylene chloride. The extracts were injected onto a liquid Chromatograph coupled with a hybrid triple quadrupole/linear ion trap mass spectrometer. Seven replicate fortifications of serum at 0.001 μg/g (1 ppb) each of penitrem A and roquefortine C gave average recoveries of 90% with 10% CV (relative standard deviation) and 97% with 3% CV, respectively. Seven replicate fortifications of urine at 0.001 μg/g (1 ppb) each of penitrem A and roquefortine C gave average recoveries of 98% with 12% CV and 100% with 6% CV, respectively. This is the first report of a positive mass spectrometric identification and quantitation of both compounds in urine and serum samples from dog intoxication cases.

Original languageEnglish (US)
Pages (from-to)4624-4629
Number of pages6
JournalAnalytical Chemistry
Volume78
Issue number13
DOIs
StatePublished - Jul 1 2006

Fingerprint

Mass spectrometers
Ions
Recovery
Methylene Chloride
roquefortine
tremortin
Liquids

ASJC Scopus subject areas

  • Analytical Chemistry

Cite this

LC-MS/MS screen for penitrem A and roquefortine C in serum and urine samples. / Tor, Elizabeth R.; Puschner, Birgit; Filigenzi, Michael S.; Tiwary, Asheesh K.; Poppenga, Robert H.

In: Analytical Chemistry, Vol. 78, No. 13, 01.07.2006, p. 4624-4629.

Research output: Contribution to journalArticle

Tor, Elizabeth R. ; Puschner, Birgit ; Filigenzi, Michael S. ; Tiwary, Asheesh K. ; Poppenga, Robert H. / LC-MS/MS screen for penitrem A and roquefortine C in serum and urine samples. In: Analytical Chemistry. 2006 ; Vol. 78, No. 13. pp. 4624-4629.
@article{c0d34833fa014340b0dfc0ce10e25700,
title = "LC-MS/MS screen for penitrem A and roquefortine C in serum and urine samples",
abstract = "A rapid LC-MS/MS method, using a triple quadrupole/linear ion trap mass spectrometer, was developed for determination of penitrem A and roquefortine C in serum and urine samples. Penitrem A and roquefortine C were extracted from samples with methylene chloride. The extracts were injected onto a liquid Chromatograph coupled with a hybrid triple quadrupole/linear ion trap mass spectrometer. Seven replicate fortifications of serum at 0.001 μg/g (1 ppb) each of penitrem A and roquefortine C gave average recoveries of 90{\%} with 10{\%} CV (relative standard deviation) and 97{\%} with 3{\%} CV, respectively. Seven replicate fortifications of urine at 0.001 μg/g (1 ppb) each of penitrem A and roquefortine C gave average recoveries of 98{\%} with 12{\%} CV and 100{\%} with 6{\%} CV, respectively. This is the first report of a positive mass spectrometric identification and quantitation of both compounds in urine and serum samples from dog intoxication cases.",
author = "Tor, {Elizabeth R.} and Birgit Puschner and Filigenzi, {Michael S.} and Tiwary, {Asheesh K.} and Poppenga, {Robert H}",
year = "2006",
month = "7",
day = "1",
doi = "10.1021/ac0601617",
language = "English (US)",
volume = "78",
pages = "4624--4629",
journal = "Analytical Chemistry",
issn = "0003-2700",
publisher = "American Chemical Society",
number = "13",

}

TY - JOUR

T1 - LC-MS/MS screen for penitrem A and roquefortine C in serum and urine samples

AU - Tor, Elizabeth R.

AU - Puschner, Birgit

AU - Filigenzi, Michael S.

AU - Tiwary, Asheesh K.

AU - Poppenga, Robert H

PY - 2006/7/1

Y1 - 2006/7/1

N2 - A rapid LC-MS/MS method, using a triple quadrupole/linear ion trap mass spectrometer, was developed for determination of penitrem A and roquefortine C in serum and urine samples. Penitrem A and roquefortine C were extracted from samples with methylene chloride. The extracts were injected onto a liquid Chromatograph coupled with a hybrid triple quadrupole/linear ion trap mass spectrometer. Seven replicate fortifications of serum at 0.001 μg/g (1 ppb) each of penitrem A and roquefortine C gave average recoveries of 90% with 10% CV (relative standard deviation) and 97% with 3% CV, respectively. Seven replicate fortifications of urine at 0.001 μg/g (1 ppb) each of penitrem A and roquefortine C gave average recoveries of 98% with 12% CV and 100% with 6% CV, respectively. This is the first report of a positive mass spectrometric identification and quantitation of both compounds in urine and serum samples from dog intoxication cases.

AB - A rapid LC-MS/MS method, using a triple quadrupole/linear ion trap mass spectrometer, was developed for determination of penitrem A and roquefortine C in serum and urine samples. Penitrem A and roquefortine C were extracted from samples with methylene chloride. The extracts were injected onto a liquid Chromatograph coupled with a hybrid triple quadrupole/linear ion trap mass spectrometer. Seven replicate fortifications of serum at 0.001 μg/g (1 ppb) each of penitrem A and roquefortine C gave average recoveries of 90% with 10% CV (relative standard deviation) and 97% with 3% CV, respectively. Seven replicate fortifications of urine at 0.001 μg/g (1 ppb) each of penitrem A and roquefortine C gave average recoveries of 98% with 12% CV and 100% with 6% CV, respectively. This is the first report of a positive mass spectrometric identification and quantitation of both compounds in urine and serum samples from dog intoxication cases.

UR - http://www.scopus.com/inward/record.url?scp=33745686146&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=33745686146&partnerID=8YFLogxK

U2 - 10.1021/ac0601617

DO - 10.1021/ac0601617

M3 - Article

C2 - 16808474

AN - SCOPUS:33745686146

VL - 78

SP - 4624

EP - 4629

JO - Analytical Chemistry

JF - Analytical Chemistry

SN - 0003-2700

IS - 13

ER -