High-resolution functional vascular assessment with ultrasound

Chih Kuang Yeh, Katherine W. Ferrara, Dustin E. Kruse

Research output: Contribution to journalArticle

23 Citations (Scopus)

Abstract

In order to improve the resolution of contrast-assisted imaging systems, we have created a high-frequency destruction/contrast replenishment imaging system with a spatial resolution of 160 μm × 160 μm. The system utilizes a 1-MHz cylindrically focused transducer for destruction and a 25-MHz spherically focused transducer for pulse/echo imaging. Speckle tracking and a clutter filter are applied across frames to remove the challenging physiologic motion artifacts that are obtained when imaging with a mechanically scanned transducer. Using a new estimation technique, flow constants proportional to absolute flow rate were estimated from B-mode time-intensity curves (TICs). The in vitro results indicate a correlation between the actual flow velocity and the estimated rate constant. In vivo images are presented showing blood perfusion in the ciliary processes and iris of the rabbit eye. The regions of interest (ROIs) from the ciliary processes yielded slower perfusion compared with the iris, as expected from vascular casts of the microcirculation in this region. Potential applications of this system include high-resolution perfusion assessment in small animals.

Original languageEnglish (US)
Pages (from-to)1263-1275
Number of pages13
JournalIEEE Transactions on Medical Imaging
Volume23
Issue number10
DOIs
StatePublished - Oct 2004

Fingerprint

Transducers
Blood Vessels
Perfusion
Ultrasonics
Iris
Imaging systems
Microcirculation
Imaging techniques
Speckle
Flow velocity
Artifacts
Rate constants
Animals
Blood
Flow rate
Rabbits

Keywords

  • Blood flow measurement
  • High-frequency ultrasound
  • Microcirculation
  • Ultrasound contrast agents

ASJC Scopus subject areas

  • Biomedical Engineering
  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology
  • Electrical and Electronic Engineering
  • Computer Science Applications
  • Computational Theory and Mathematics

Cite this

High-resolution functional vascular assessment with ultrasound. / Yeh, Chih Kuang; Ferrara, Katherine W.; Kruse, Dustin E.

In: IEEE Transactions on Medical Imaging, Vol. 23, No. 10, 10.2004, p. 1263-1275.

Research output: Contribution to journalArticle

Yeh, Chih Kuang ; Ferrara, Katherine W. ; Kruse, Dustin E. / High-resolution functional vascular assessment with ultrasound. In: IEEE Transactions on Medical Imaging. 2004 ; Vol. 23, No. 10. pp. 1263-1275.
@article{b95f6d97987b4ac180ca621e116dec98,
title = "High-resolution functional vascular assessment with ultrasound",
abstract = "In order to improve the resolution of contrast-assisted imaging systems, we have created a high-frequency destruction/contrast replenishment imaging system with a spatial resolution of 160 μm × 160 μm. The system utilizes a 1-MHz cylindrically focused transducer for destruction and a 25-MHz spherically focused transducer for pulse/echo imaging. Speckle tracking and a clutter filter are applied across frames to remove the challenging physiologic motion artifacts that are obtained when imaging with a mechanically scanned transducer. Using a new estimation technique, flow constants proportional to absolute flow rate were estimated from B-mode time-intensity curves (TICs). The in vitro results indicate a correlation between the actual flow velocity and the estimated rate constant. In vivo images are presented showing blood perfusion in the ciliary processes and iris of the rabbit eye. The regions of interest (ROIs) from the ciliary processes yielded slower perfusion compared with the iris, as expected from vascular casts of the microcirculation in this region. Potential applications of this system include high-resolution perfusion assessment in small animals.",
keywords = "Blood flow measurement, High-frequency ultrasound, Microcirculation, Ultrasound contrast agents",
author = "Yeh, {Chih Kuang} and Ferrara, {Katherine W.} and Kruse, {Dustin E.}",
year = "2004",
month = "10",
doi = "10.1109/TMI.2004.834614",
language = "English (US)",
volume = "23",
pages = "1263--1275",
journal = "IEEE Transactions on Medical Imaging",
issn = "0278-0062",
publisher = "Institute of Electrical and Electronics Engineers Inc.",
number = "10",

}

TY - JOUR

T1 - High-resolution functional vascular assessment with ultrasound

AU - Yeh, Chih Kuang

AU - Ferrara, Katherine W.

AU - Kruse, Dustin E.

PY - 2004/10

Y1 - 2004/10

N2 - In order to improve the resolution of contrast-assisted imaging systems, we have created a high-frequency destruction/contrast replenishment imaging system with a spatial resolution of 160 μm × 160 μm. The system utilizes a 1-MHz cylindrically focused transducer for destruction and a 25-MHz spherically focused transducer for pulse/echo imaging. Speckle tracking and a clutter filter are applied across frames to remove the challenging physiologic motion artifacts that are obtained when imaging with a mechanically scanned transducer. Using a new estimation technique, flow constants proportional to absolute flow rate were estimated from B-mode time-intensity curves (TICs). The in vitro results indicate a correlation between the actual flow velocity and the estimated rate constant. In vivo images are presented showing blood perfusion in the ciliary processes and iris of the rabbit eye. The regions of interest (ROIs) from the ciliary processes yielded slower perfusion compared with the iris, as expected from vascular casts of the microcirculation in this region. Potential applications of this system include high-resolution perfusion assessment in small animals.

AB - In order to improve the resolution of contrast-assisted imaging systems, we have created a high-frequency destruction/contrast replenishment imaging system with a spatial resolution of 160 μm × 160 μm. The system utilizes a 1-MHz cylindrically focused transducer for destruction and a 25-MHz spherically focused transducer for pulse/echo imaging. Speckle tracking and a clutter filter are applied across frames to remove the challenging physiologic motion artifacts that are obtained when imaging with a mechanically scanned transducer. Using a new estimation technique, flow constants proportional to absolute flow rate were estimated from B-mode time-intensity curves (TICs). The in vitro results indicate a correlation between the actual flow velocity and the estimated rate constant. In vivo images are presented showing blood perfusion in the ciliary processes and iris of the rabbit eye. The regions of interest (ROIs) from the ciliary processes yielded slower perfusion compared with the iris, as expected from vascular casts of the microcirculation in this region. Potential applications of this system include high-resolution perfusion assessment in small animals.

KW - Blood flow measurement

KW - High-frequency ultrasound

KW - Microcirculation

KW - Ultrasound contrast agents

UR - http://www.scopus.com/inward/record.url?scp=6344250735&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=6344250735&partnerID=8YFLogxK

U2 - 10.1109/TMI.2004.834614

DO - 10.1109/TMI.2004.834614

M3 - Article

VL - 23

SP - 1263

EP - 1275

JO - IEEE Transactions on Medical Imaging

JF - IEEE Transactions on Medical Imaging

SN - 0278-0062

IS - 10

ER -