Film dosimetry in the peripheral region using multiple sensitometric curves

Robin L Stern, John Kurylo, Christine Hartmann Siantar, Joerg Lehmann, Zelanna Goldberg

Research output: Contribution to journalArticle

7 Citations (Scopus)

Abstract

We present a method for applying film dosimetry to the peripheral region utilizing multiple sensitometric curves. There are many instances when the dose to the peripheral region outside the field edges is of clinical and/or research interest. Published peripheral dose data may be insufficient if detailed dose modeling is required, and in those cases measurements must be performed. Film dosimetry is an attractive approach for dose measurement in the peripheral region because it integrates dose, overcoming the low-dose-rate problem, and is time efficient, as it acquires an entire plan of data in a single exposure. However, film response increases at energies below approximately 300 keV. As the scattered photon spectrum changes with distance from the field edge, this increased film sensitivity causes changes in the film response along profiles perpendicular to the field edge. A single sensitometric curve is therefore no longer sufficient for accurate conversion of the optical density to dose. Our new method uses multiple sensitometric curves defined at increasing distances from the field edge. To convert an optical density profile, the dose at each point in the profile is defined as a linear combination of the doses calculated using the two sensitometric curves that bracket the point of interest. A single set of sensitometric curves derived at one field size and source-to-surface distance (SSD) can be applied to density profiles for other field sizes and SSDs. We verified our new method by comparison to ion chamber measurements using three different types of film. Agreement with chamber measurements was within 7%, or less than 2 mm in regions of high gradient, over a wide range of field sizes and SSDs.

Original languageEnglish (US)
Pages (from-to)327-332
Number of pages6
JournalMedical Physics
Volume31
Issue number2
DOIs
StatePublished - Feb 2004

Fingerprint

Film Dosimetry
Silver Sulfadiazine
Photons
Ions
Research

Keywords

  • Film dosimetry
  • Peripheral dose
  • Radiation dosimetry

ASJC Scopus subject areas

  • Biophysics

Cite this

Film dosimetry in the peripheral region using multiple sensitometric curves. / Stern, Robin L; Kurylo, John; Siantar, Christine Hartmann; Lehmann, Joerg; Goldberg, Zelanna.

In: Medical Physics, Vol. 31, No. 2, 02.2004, p. 327-332.

Research output: Contribution to journalArticle

Stern, RL, Kurylo, J, Siantar, CH, Lehmann, J & Goldberg, Z 2004, 'Film dosimetry in the peripheral region using multiple sensitometric curves', Medical Physics, vol. 31, no. 2, pp. 327-332. https://doi.org/10.1118/1.1639126
Stern, Robin L ; Kurylo, John ; Siantar, Christine Hartmann ; Lehmann, Joerg ; Goldberg, Zelanna. / Film dosimetry in the peripheral region using multiple sensitometric curves. In: Medical Physics. 2004 ; Vol. 31, No. 2. pp. 327-332.
@article{b0bba04955164fa98735dcd945ae8960,
title = "Film dosimetry in the peripheral region using multiple sensitometric curves",
abstract = "We present a method for applying film dosimetry to the peripheral region utilizing multiple sensitometric curves. There are many instances when the dose to the peripheral region outside the field edges is of clinical and/or research interest. Published peripheral dose data may be insufficient if detailed dose modeling is required, and in those cases measurements must be performed. Film dosimetry is an attractive approach for dose measurement in the peripheral region because it integrates dose, overcoming the low-dose-rate problem, and is time efficient, as it acquires an entire plan of data in a single exposure. However, film response increases at energies below approximately 300 keV. As the scattered photon spectrum changes with distance from the field edge, this increased film sensitivity causes changes in the film response along profiles perpendicular to the field edge. A single sensitometric curve is therefore no longer sufficient for accurate conversion of the optical density to dose. Our new method uses multiple sensitometric curves defined at increasing distances from the field edge. To convert an optical density profile, the dose at each point in the profile is defined as a linear combination of the doses calculated using the two sensitometric curves that bracket the point of interest. A single set of sensitometric curves derived at one field size and source-to-surface distance (SSD) can be applied to density profiles for other field sizes and SSDs. We verified our new method by comparison to ion chamber measurements using three different types of film. Agreement with chamber measurements was within 7{\%}, or less than 2 mm in regions of high gradient, over a wide range of field sizes and SSDs.",
keywords = "Film dosimetry, Peripheral dose, Radiation dosimetry",
author = "Stern, {Robin L} and John Kurylo and Siantar, {Christine Hartmann} and Joerg Lehmann and Zelanna Goldberg",
year = "2004",
month = "2",
doi = "10.1118/1.1639126",
language = "English (US)",
volume = "31",
pages = "327--332",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "2",

}

TY - JOUR

T1 - Film dosimetry in the peripheral region using multiple sensitometric curves

AU - Stern, Robin L

AU - Kurylo, John

AU - Siantar, Christine Hartmann

AU - Lehmann, Joerg

AU - Goldberg, Zelanna

PY - 2004/2

Y1 - 2004/2

N2 - We present a method for applying film dosimetry to the peripheral region utilizing multiple sensitometric curves. There are many instances when the dose to the peripheral region outside the field edges is of clinical and/or research interest. Published peripheral dose data may be insufficient if detailed dose modeling is required, and in those cases measurements must be performed. Film dosimetry is an attractive approach for dose measurement in the peripheral region because it integrates dose, overcoming the low-dose-rate problem, and is time efficient, as it acquires an entire plan of data in a single exposure. However, film response increases at energies below approximately 300 keV. As the scattered photon spectrum changes with distance from the field edge, this increased film sensitivity causes changes in the film response along profiles perpendicular to the field edge. A single sensitometric curve is therefore no longer sufficient for accurate conversion of the optical density to dose. Our new method uses multiple sensitometric curves defined at increasing distances from the field edge. To convert an optical density profile, the dose at each point in the profile is defined as a linear combination of the doses calculated using the two sensitometric curves that bracket the point of interest. A single set of sensitometric curves derived at one field size and source-to-surface distance (SSD) can be applied to density profiles for other field sizes and SSDs. We verified our new method by comparison to ion chamber measurements using three different types of film. Agreement with chamber measurements was within 7%, or less than 2 mm in regions of high gradient, over a wide range of field sizes and SSDs.

AB - We present a method for applying film dosimetry to the peripheral region utilizing multiple sensitometric curves. There are many instances when the dose to the peripheral region outside the field edges is of clinical and/or research interest. Published peripheral dose data may be insufficient if detailed dose modeling is required, and in those cases measurements must be performed. Film dosimetry is an attractive approach for dose measurement in the peripheral region because it integrates dose, overcoming the low-dose-rate problem, and is time efficient, as it acquires an entire plan of data in a single exposure. However, film response increases at energies below approximately 300 keV. As the scattered photon spectrum changes with distance from the field edge, this increased film sensitivity causes changes in the film response along profiles perpendicular to the field edge. A single sensitometric curve is therefore no longer sufficient for accurate conversion of the optical density to dose. Our new method uses multiple sensitometric curves defined at increasing distances from the field edge. To convert an optical density profile, the dose at each point in the profile is defined as a linear combination of the doses calculated using the two sensitometric curves that bracket the point of interest. A single set of sensitometric curves derived at one field size and source-to-surface distance (SSD) can be applied to density profiles for other field sizes and SSDs. We verified our new method by comparison to ion chamber measurements using three different types of film. Agreement with chamber measurements was within 7%, or less than 2 mm in regions of high gradient, over a wide range of field sizes and SSDs.

KW - Film dosimetry

KW - Peripheral dose

KW - Radiation dosimetry

UR - http://www.scopus.com/inward/record.url?scp=1242338922&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=1242338922&partnerID=8YFLogxK

U2 - 10.1118/1.1639126

DO - 10.1118/1.1639126

M3 - Article

C2 - 15000618

AN - SCOPUS:1242338922

VL - 31

SP - 327

EP - 332

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 2

ER -