Fabrication and characterization of a 0.5-mm lutetium oxyorthosilicate detector array for high-resolution PET applications

Jennifer R. Stickel, Jinyi Qi, Simon R Cherry

Research output: Contribution to journalArticle

54 Citations (Scopus)

Abstract

With the increasing use of in vivo imaging in mouse models of disease, there are many interesting applications that demand imaging of organs and tissues with submillimeter resolution. Though there are other contributing factors, the spatial resolution in small-animal PET is still largely determined by the detector pixel dimensions. Methods: In this work, a pair of lutetium oxyorthosilicate (LSO) arrays with 0.5-mm pixelswas coupled tomultichannel photomultiplier tubes and evaluated for use as high-resolution PET detectors. Results: Flood histograms demonstrated that most crystals were clearly identifiable. Energy resolution varied from 22% to 38%. The coincidence timing resolution was 1.42-ns full width at half maximum (FWHM). The intrinsic spatial resolution was 0.68-mm FWHM as measured with a 30-gauge needle filled with 18F. The improvement in spatial resolution in a tomographic setting is demonstrated using images of a line source phantom reconstructed with filtered backprojection and compared with images obtained from 2 dedicated small-animal PET scanners. Finally, a projection image of the mouse foot is shown to demonstrate the application of these 0.5-mm LSO detectors to a biologic task. Conclusion: A pair of highly pixelated LSO detections has been constructed and characterized for use as high-spatial-resolution PET detectors. It appears that small-animal PET systems capable of a FWHM spatial resolution of 600 μm or less are feasible and should be pursued.

Original languageEnglish (US)
Pages (from-to)115-121
Number of pages7
JournalJournal of Nuclear Medicine
Volume48
Issue number1
StatePublished - Jan 1 2007

Fingerprint

Needles
Foot
lutetium orthosilicate

Keywords

  • High spatial resolution
  • Instrumentation
  • Lutetium oxyorthosilicate
  • Small-animal PET

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology

Cite this

Fabrication and characterization of a 0.5-mm lutetium oxyorthosilicate detector array for high-resolution PET applications. / Stickel, Jennifer R.; Qi, Jinyi; Cherry, Simon R.

In: Journal of Nuclear Medicine, Vol. 48, No. 1, 01.01.2007, p. 115-121.

Research output: Contribution to journalArticle

@article{adc4535480464eb4b40a161ef2f4114c,
title = "Fabrication and characterization of a 0.5-mm lutetium oxyorthosilicate detector array for high-resolution PET applications",
abstract = "With the increasing use of in vivo imaging in mouse models of disease, there are many interesting applications that demand imaging of organs and tissues with submillimeter resolution. Though there are other contributing factors, the spatial resolution in small-animal PET is still largely determined by the detector pixel dimensions. Methods: In this work, a pair of lutetium oxyorthosilicate (LSO) arrays with 0.5-mm pixelswas coupled tomultichannel photomultiplier tubes and evaluated for use as high-resolution PET detectors. Results: Flood histograms demonstrated that most crystals were clearly identifiable. Energy resolution varied from 22{\%} to 38{\%}. The coincidence timing resolution was 1.42-ns full width at half maximum (FWHM). The intrinsic spatial resolution was 0.68-mm FWHM as measured with a 30-gauge needle filled with 18F. The improvement in spatial resolution in a tomographic setting is demonstrated using images of a line source phantom reconstructed with filtered backprojection and compared with images obtained from 2 dedicated small-animal PET scanners. Finally, a projection image of the mouse foot is shown to demonstrate the application of these 0.5-mm LSO detectors to a biologic task. Conclusion: A pair of highly pixelated LSO detections has been constructed and characterized for use as high-spatial-resolution PET detectors. It appears that small-animal PET systems capable of a FWHM spatial resolution of 600 μm or less are feasible and should be pursued.",
keywords = "High spatial resolution, Instrumentation, Lutetium oxyorthosilicate, Small-animal PET",
author = "Stickel, {Jennifer R.} and Jinyi Qi and Cherry, {Simon R}",
year = "2007",
month = "1",
day = "1",
language = "English (US)",
volume = "48",
pages = "115--121",
journal = "Journal of Nuclear Medicine",
issn = "0161-5505",
publisher = "Society of Nuclear Medicine Inc.",
number = "1",

}

TY - JOUR

T1 - Fabrication and characterization of a 0.5-mm lutetium oxyorthosilicate detector array for high-resolution PET applications

AU - Stickel, Jennifer R.

AU - Qi, Jinyi

AU - Cherry, Simon R

PY - 2007/1/1

Y1 - 2007/1/1

N2 - With the increasing use of in vivo imaging in mouse models of disease, there are many interesting applications that demand imaging of organs and tissues with submillimeter resolution. Though there are other contributing factors, the spatial resolution in small-animal PET is still largely determined by the detector pixel dimensions. Methods: In this work, a pair of lutetium oxyorthosilicate (LSO) arrays with 0.5-mm pixelswas coupled tomultichannel photomultiplier tubes and evaluated for use as high-resolution PET detectors. Results: Flood histograms demonstrated that most crystals were clearly identifiable. Energy resolution varied from 22% to 38%. The coincidence timing resolution was 1.42-ns full width at half maximum (FWHM). The intrinsic spatial resolution was 0.68-mm FWHM as measured with a 30-gauge needle filled with 18F. The improvement in spatial resolution in a tomographic setting is demonstrated using images of a line source phantom reconstructed with filtered backprojection and compared with images obtained from 2 dedicated small-animal PET scanners. Finally, a projection image of the mouse foot is shown to demonstrate the application of these 0.5-mm LSO detectors to a biologic task. Conclusion: A pair of highly pixelated LSO detections has been constructed and characterized for use as high-spatial-resolution PET detectors. It appears that small-animal PET systems capable of a FWHM spatial resolution of 600 μm or less are feasible and should be pursued.

AB - With the increasing use of in vivo imaging in mouse models of disease, there are many interesting applications that demand imaging of organs and tissues with submillimeter resolution. Though there are other contributing factors, the spatial resolution in small-animal PET is still largely determined by the detector pixel dimensions. Methods: In this work, a pair of lutetium oxyorthosilicate (LSO) arrays with 0.5-mm pixelswas coupled tomultichannel photomultiplier tubes and evaluated for use as high-resolution PET detectors. Results: Flood histograms demonstrated that most crystals were clearly identifiable. Energy resolution varied from 22% to 38%. The coincidence timing resolution was 1.42-ns full width at half maximum (FWHM). The intrinsic spatial resolution was 0.68-mm FWHM as measured with a 30-gauge needle filled with 18F. The improvement in spatial resolution in a tomographic setting is demonstrated using images of a line source phantom reconstructed with filtered backprojection and compared with images obtained from 2 dedicated small-animal PET scanners. Finally, a projection image of the mouse foot is shown to demonstrate the application of these 0.5-mm LSO detectors to a biologic task. Conclusion: A pair of highly pixelated LSO detections has been constructed and characterized for use as high-spatial-resolution PET detectors. It appears that small-animal PET systems capable of a FWHM spatial resolution of 600 μm or less are feasible and should be pursued.

KW - High spatial resolution

KW - Instrumentation

KW - Lutetium oxyorthosilicate

KW - Small-animal PET

UR - http://www.scopus.com/inward/record.url?scp=33847264981&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=33847264981&partnerID=8YFLogxK

M3 - Article

C2 - 17204707

AN - SCOPUS:33847264981

VL - 48

SP - 115

EP - 121

JO - Journal of Nuclear Medicine

JF - Journal of Nuclear Medicine

SN - 0161-5505

IS - 1

ER -