External fixator clamp reuse degrades clamp mechanical performance

Robert S. Gilley, David P. Beason, David M. Snyder, Raymond C. Boston, Alex Radin, Amy S. Kapatkin, Gail K. Smith, Louis J. Soslowsky

Research output: Contribution to journalArticle

8 Scopus citations

Abstract

Objective To determine the effects of clamp reuse for the Kirschner-Ehmer (KE); Securos; and the IMEX-SK clamp. Study Design Experimental bench test of mechanical properties. Methods Specially designed fixtures were used to mechanically test 18 clamps of each type with respect to 6 mechanical variables: fixator pin slippage, connecting bar slippage, fixator pin rotation, connecting bar rotation, and clockwise and counterclockwise clamp-bolt axis pivot. Each clamp was tested 6 times for each variable at 7.68 Nm of clamp-bolt tightening torque. Results were compared using repeated measures ANOVA. Results For the IMEX-SK clamp, a significant degradation in the force required to cause slippage was found for connecting bar slippage and fixator pin rotation; however, this clamp also had a significant increase in the force to initiate slippage for clockwise clamp-bolt axis pivot with reuse. The Securos clamp had significant degradation in connecting bar slippage, connecting bar rotation, and fixator pin rotation whereas the KE clamp had significant degradation in connecting bar slippage only. Conclusions All 3 external fixator clamp types degraded in 1 or more movement variables in their ability to resist motion with reuse. The IMEX and Securos clamps were more subject to degradation than the KE clamp and this may have clinical importance for fixator composite rigidity. Clinical Relevance Fracture-reduction stability is related to the ability to resist motion within a clamp. The unpredictable nature of degradation we found cautions against repeated use. Clinicians should consider reuse of external fixator clamps with the knowledge that repeated use degrades clamp mechanical performance.

Original languageEnglish (US)
Pages (from-to)530-536
Number of pages7
JournalVeterinary Surgery
Volume38
Issue number4
DOIs
StatePublished - Jun 2009

ASJC Scopus subject areas

  • veterinary(all)

Fingerprint Dive into the research topics of 'External fixator clamp reuse degrades clamp mechanical performance'. Together they form a unique fingerprint.

  • Cite this

    Gilley, R. S., Beason, D. P., Snyder, D. M., Boston, R. C., Radin, A., Kapatkin, A. S., Smith, G. K., & Soslowsky, L. J. (2009). External fixator clamp reuse degrades clamp mechanical performance. Veterinary Surgery, 38(4), 530-536. https://doi.org/10.1111/j.1532-950X.2009.00521.x