Evaluation of x-ray scatter properties in a dedicated cone-beam breast CT scanner

Alexander L C Kwan, John M Boone, Nikula Shah

Research output: Contribution to journalArticle

92 Citations (Scopus)

Abstract

The magnitude of scatter contamination on a first-generation prototype breast computed tomography (CT) scanner was evaluated using the scatter-to-primary ratio (SPR) metric. The SPR was measured and characterized over a wide range of parameters relevant to breast CT imaging, including x-ray beam energy, breast diameter, breast composition, isocenter-to-detector distance, collimated slot thickness, and grid ratio. The results demonstrated that in the absence of scatter reduction techniques, the SPR levels for the average breast (e.g., 14 cm diameter 50/50 composition cylindrical phantom) are quite high (∼0.5 at the center of the phantom for 80 kVp in true cone-beam CT geometry), and increases as the diameter of the phantom is increased (to ∼1.0 at the center of a 18 cm diameter 50/50 phantom). The x-ray beam energy and the phantom compositions had only minimal impact on the measured SPR. When an ideal bowtie filter was used, the SPRs at the central axis of the 14 and 18 cm cylindrical phantoms were reduced while the SPRs at the edge of the phantoms were increased. Lastly, collimation in the vertical direction had a significant impact on the SPRs at the central axis of the phantoms. These high SPR levels might lead to cupping artifacts and increased noise in the reconstructed CT images, and this suggests that efficient scatter rejection and/or correction techniques may be required to improve the quality and accuracy of cone beam CT images.

Original languageEnglish (US)
Pages (from-to)2967-2975
Number of pages9
JournalMedical Physics
Volume32
Issue number9
DOIs
StatePublished - Sep 2005

Fingerprint

Cone-Beam Computed Tomography
Breast
X-Rays
Tomography
X-Ray Computed Tomography Scanners
Artifacts
Noise

Keywords

  • Breast imaging
  • Computed tomography
  • Cone-beam
  • Mammography
  • Scatter-to-primary ratio
  • SPR
  • X-ray

ASJC Scopus subject areas

  • Biophysics

Cite this

Evaluation of x-ray scatter properties in a dedicated cone-beam breast CT scanner. / Kwan, Alexander L C; Boone, John M; Shah, Nikula.

In: Medical Physics, Vol. 32, No. 9, 09.2005, p. 2967-2975.

Research output: Contribution to journalArticle

Kwan, Alexander L C ; Boone, John M ; Shah, Nikula. / Evaluation of x-ray scatter properties in a dedicated cone-beam breast CT scanner. In: Medical Physics. 2005 ; Vol. 32, No. 9. pp. 2967-2975.
@article{edef886280b44e28bf6acde91752f275,
title = "Evaluation of x-ray scatter properties in a dedicated cone-beam breast CT scanner",
abstract = "The magnitude of scatter contamination on a first-generation prototype breast computed tomography (CT) scanner was evaluated using the scatter-to-primary ratio (SPR) metric. The SPR was measured and characterized over a wide range of parameters relevant to breast CT imaging, including x-ray beam energy, breast diameter, breast composition, isocenter-to-detector distance, collimated slot thickness, and grid ratio. The results demonstrated that in the absence of scatter reduction techniques, the SPR levels for the average breast (e.g., 14 cm diameter 50/50 composition cylindrical phantom) are quite high (∼0.5 at the center of the phantom for 80 kVp in true cone-beam CT geometry), and increases as the diameter of the phantom is increased (to ∼1.0 at the center of a 18 cm diameter 50/50 phantom). The x-ray beam energy and the phantom compositions had only minimal impact on the measured SPR. When an ideal bowtie filter was used, the SPRs at the central axis of the 14 and 18 cm cylindrical phantoms were reduced while the SPRs at the edge of the phantoms were increased. Lastly, collimation in the vertical direction had a significant impact on the SPRs at the central axis of the phantoms. These high SPR levels might lead to cupping artifacts and increased noise in the reconstructed CT images, and this suggests that efficient scatter rejection and/or correction techniques may be required to improve the quality and accuracy of cone beam CT images.",
keywords = "Breast imaging, Computed tomography, Cone-beam, Mammography, Scatter-to-primary ratio, SPR, X-ray",
author = "Kwan, {Alexander L C} and Boone, {John M} and Nikula Shah",
year = "2005",
month = "9",
doi = "10.1118/1.1954908",
language = "English (US)",
volume = "32",
pages = "2967--2975",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "9",

}

TY - JOUR

T1 - Evaluation of x-ray scatter properties in a dedicated cone-beam breast CT scanner

AU - Kwan, Alexander L C

AU - Boone, John M

AU - Shah, Nikula

PY - 2005/9

Y1 - 2005/9

N2 - The magnitude of scatter contamination on a first-generation prototype breast computed tomography (CT) scanner was evaluated using the scatter-to-primary ratio (SPR) metric. The SPR was measured and characterized over a wide range of parameters relevant to breast CT imaging, including x-ray beam energy, breast diameter, breast composition, isocenter-to-detector distance, collimated slot thickness, and grid ratio. The results demonstrated that in the absence of scatter reduction techniques, the SPR levels for the average breast (e.g., 14 cm diameter 50/50 composition cylindrical phantom) are quite high (∼0.5 at the center of the phantom for 80 kVp in true cone-beam CT geometry), and increases as the diameter of the phantom is increased (to ∼1.0 at the center of a 18 cm diameter 50/50 phantom). The x-ray beam energy and the phantom compositions had only minimal impact on the measured SPR. When an ideal bowtie filter was used, the SPRs at the central axis of the 14 and 18 cm cylindrical phantoms were reduced while the SPRs at the edge of the phantoms were increased. Lastly, collimation in the vertical direction had a significant impact on the SPRs at the central axis of the phantoms. These high SPR levels might lead to cupping artifacts and increased noise in the reconstructed CT images, and this suggests that efficient scatter rejection and/or correction techniques may be required to improve the quality and accuracy of cone beam CT images.

AB - The magnitude of scatter contamination on a first-generation prototype breast computed tomography (CT) scanner was evaluated using the scatter-to-primary ratio (SPR) metric. The SPR was measured and characterized over a wide range of parameters relevant to breast CT imaging, including x-ray beam energy, breast diameter, breast composition, isocenter-to-detector distance, collimated slot thickness, and grid ratio. The results demonstrated that in the absence of scatter reduction techniques, the SPR levels for the average breast (e.g., 14 cm diameter 50/50 composition cylindrical phantom) are quite high (∼0.5 at the center of the phantom for 80 kVp in true cone-beam CT geometry), and increases as the diameter of the phantom is increased (to ∼1.0 at the center of a 18 cm diameter 50/50 phantom). The x-ray beam energy and the phantom compositions had only minimal impact on the measured SPR. When an ideal bowtie filter was used, the SPRs at the central axis of the 14 and 18 cm cylindrical phantoms were reduced while the SPRs at the edge of the phantoms were increased. Lastly, collimation in the vertical direction had a significant impact on the SPRs at the central axis of the phantoms. These high SPR levels might lead to cupping artifacts and increased noise in the reconstructed CT images, and this suggests that efficient scatter rejection and/or correction techniques may be required to improve the quality and accuracy of cone beam CT images.

KW - Breast imaging

KW - Computed tomography

KW - Cone-beam

KW - Mammography

KW - Scatter-to-primary ratio

KW - SPR

KW - X-ray

UR - http://www.scopus.com/inward/record.url?scp=25644435753&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=25644435753&partnerID=8YFLogxK

U2 - 10.1118/1.1954908

DO - 10.1118/1.1954908

M3 - Article

C2 - 16266111

AN - SCOPUS:25644435753

VL - 32

SP - 2967

EP - 2975

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 9

ER -