Effect of phantom voxelization in CT simulations

Andrew L. Goertzen, Freek J. Beekman, Simon R Cherry

Research output: Contribution to journalArticle

23 Citations (Scopus)

Abstract

In computer simulations of x-ray CT systems one can either use continuous geometrical descriptions for phantoms or a voxelized representation. The voxelized approach allows arbitrary phantoms to be defined without being confined to geometrical shapes. The disadvantage of the voxelized approach is that inherent errors are introduced due to the phantom voxelization. To study effects of phantom discretization, analytical CT simulations were run for a fan-beam geometry with phantom voxel sizes ranging from 0.0625 to 2 times the reconstructed pixel size and noise levels corresponding to 103-107 photons per detector pixel prior to attenuation. The number of rays traced per detector element was varied from 1 to 16. Differences in the filtered backprojection images caused by changing the phantom matrix sizes and number of rays traced were assessed by calculating the difference between reconstructions based on the finest matrix and coarser matrix simulations. In noise free simulations, all phantom matrix sizes produced a measurable difference in comparison with the finest phantom matrix used. When even a small amount of noise was added to the projection data, the differences due to the phantom discretization were masked by the noise, and in all cases there was almost no improvement by using a phantom matrix that was more than twice as fine as the reconstruction matrix. No substantial improvement was achieved by tracing more than 4 rays per detector pixel.

Original languageEnglish (US)
Pages (from-to)492-498
Number of pages7
JournalMedical Physics
Volume29
Issue number4
DOIs
StatePublished - 2002

Fingerprint

Noise
Photons
Computer Simulation
X-Rays

Keywords

  • Artifacts
  • Computed tomography
  • CT image quality
  • Phantom study
  • Simulation

ASJC Scopus subject areas

  • Biophysics

Cite this

Effect of phantom voxelization in CT simulations. / Goertzen, Andrew L.; Beekman, Freek J.; Cherry, Simon R.

In: Medical Physics, Vol. 29, No. 4, 2002, p. 492-498.

Research output: Contribution to journalArticle

Goertzen, Andrew L. ; Beekman, Freek J. ; Cherry, Simon R. / Effect of phantom voxelization in CT simulations. In: Medical Physics. 2002 ; Vol. 29, No. 4. pp. 492-498.
@article{fb76f5d1899f4d7f8215b975833ce47b,
title = "Effect of phantom voxelization in CT simulations",
abstract = "In computer simulations of x-ray CT systems one can either use continuous geometrical descriptions for phantoms or a voxelized representation. The voxelized approach allows arbitrary phantoms to be defined without being confined to geometrical shapes. The disadvantage of the voxelized approach is that inherent errors are introduced due to the phantom voxelization. To study effects of phantom discretization, analytical CT simulations were run for a fan-beam geometry with phantom voxel sizes ranging from 0.0625 to 2 times the reconstructed pixel size and noise levels corresponding to 103-107 photons per detector pixel prior to attenuation. The number of rays traced per detector element was varied from 1 to 16. Differences in the filtered backprojection images caused by changing the phantom matrix sizes and number of rays traced were assessed by calculating the difference between reconstructions based on the finest matrix and coarser matrix simulations. In noise free simulations, all phantom matrix sizes produced a measurable difference in comparison with the finest phantom matrix used. When even a small amount of noise was added to the projection data, the differences due to the phantom discretization were masked by the noise, and in all cases there was almost no improvement by using a phantom matrix that was more than twice as fine as the reconstruction matrix. No substantial improvement was achieved by tracing more than 4 rays per detector pixel.",
keywords = "Artifacts, Computed tomography, CT image quality, Phantom study, Simulation",
author = "Goertzen, {Andrew L.} and Beekman, {Freek J.} and Cherry, {Simon R}",
year = "2002",
doi = "10.1118/1.1461841",
language = "English (US)",
volume = "29",
pages = "492--498",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "4",

}

TY - JOUR

T1 - Effect of phantom voxelization in CT simulations

AU - Goertzen, Andrew L.

AU - Beekman, Freek J.

AU - Cherry, Simon R

PY - 2002

Y1 - 2002

N2 - In computer simulations of x-ray CT systems one can either use continuous geometrical descriptions for phantoms or a voxelized representation. The voxelized approach allows arbitrary phantoms to be defined without being confined to geometrical shapes. The disadvantage of the voxelized approach is that inherent errors are introduced due to the phantom voxelization. To study effects of phantom discretization, analytical CT simulations were run for a fan-beam geometry with phantom voxel sizes ranging from 0.0625 to 2 times the reconstructed pixel size and noise levels corresponding to 103-107 photons per detector pixel prior to attenuation. The number of rays traced per detector element was varied from 1 to 16. Differences in the filtered backprojection images caused by changing the phantom matrix sizes and number of rays traced were assessed by calculating the difference between reconstructions based on the finest matrix and coarser matrix simulations. In noise free simulations, all phantom matrix sizes produced a measurable difference in comparison with the finest phantom matrix used. When even a small amount of noise was added to the projection data, the differences due to the phantom discretization were masked by the noise, and in all cases there was almost no improvement by using a phantom matrix that was more than twice as fine as the reconstruction matrix. No substantial improvement was achieved by tracing more than 4 rays per detector pixel.

AB - In computer simulations of x-ray CT systems one can either use continuous geometrical descriptions for phantoms or a voxelized representation. The voxelized approach allows arbitrary phantoms to be defined without being confined to geometrical shapes. The disadvantage of the voxelized approach is that inherent errors are introduced due to the phantom voxelization. To study effects of phantom discretization, analytical CT simulations were run for a fan-beam geometry with phantom voxel sizes ranging from 0.0625 to 2 times the reconstructed pixel size and noise levels corresponding to 103-107 photons per detector pixel prior to attenuation. The number of rays traced per detector element was varied from 1 to 16. Differences in the filtered backprojection images caused by changing the phantom matrix sizes and number of rays traced were assessed by calculating the difference between reconstructions based on the finest matrix and coarser matrix simulations. In noise free simulations, all phantom matrix sizes produced a measurable difference in comparison with the finest phantom matrix used. When even a small amount of noise was added to the projection data, the differences due to the phantom discretization were masked by the noise, and in all cases there was almost no improvement by using a phantom matrix that was more than twice as fine as the reconstruction matrix. No substantial improvement was achieved by tracing more than 4 rays per detector pixel.

KW - Artifacts

KW - Computed tomography

KW - CT image quality

KW - Phantom study

KW - Simulation

UR - http://www.scopus.com/inward/record.url?scp=0036120688&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0036120688&partnerID=8YFLogxK

U2 - 10.1118/1.1461841

DO - 10.1118/1.1461841

M3 - Article

C2 - 11991120

AN - SCOPUS:0036120688

VL - 29

SP - 492

EP - 498

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 4

ER -