Abstract

A microfabricated corona ionizer is developed for miniaturized air particle monitoring instruments. Peek's breakdown criterion and Warburg's Law are used to calculate the required electrode geometries. Single mask electroplating is employed as the microfabrication technique. Devices with discharge pin radius of 20 μm and interelectrode spacing of 1.4, 1.8, 2.0, and 2.2 mm are fabricated and tested. The inception voltages in the range from 1.4 to 2 kV are achieved without substrate breakdown. It consumes relatively low power at 150 mW or less. Corona charging currents of more than 50 μA were achieved before streamers break down. The experimental results also demonstrated the nondependency of the inception voltage on the pin's radius of curvature so long as it is significantly smaller than the interelectrode spacing. There is only one value of inception voltage for each given interelectrode spacing (1.4 to 2.2 mm) with the pin's radius of curvature ranging from 10 to 200 μm. This implies that strict dimensional control is not required in the microfabrication process in order to ensure the same inception voltage for each device.

Original languageEnglish (US)
Pages (from-to)115-123
Number of pages9
JournalJournal of Microelectromechanical Systems
Volume17
Issue number1
DOIs
StatePublished - Feb 2008

Fingerprint

Fabrication
Microfabrication
Testing
Electric potential
Electroplating
Masks
Electrodes
Geometry
Monitoring
Substrates
Air

Keywords

  • Avalanche
  • Corona
  • Discharge
  • Electroplating
  • Inception
  • Ionizer
  • Microfabricated
  • Plasma
  • Streamer

ASJC Scopus subject areas

  • Electrical and Electronic Engineering
  • Mechanical Engineering

Cite this

Design, fabrication, and testing of a microfabricated Corona Ionizer. / Chua, Beelee; Wexler, Anthony S.; Tien, Norman C.; Niemeier, Debbie A.; Holmen, Britt A.

In: Journal of Microelectromechanical Systems, Vol. 17, No. 1, 02.2008, p. 115-123.

Research output: Contribution to journalArticle

Chua, Beelee ; Wexler, Anthony S. ; Tien, Norman C. ; Niemeier, Debbie A. ; Holmen, Britt A. / Design, fabrication, and testing of a microfabricated Corona Ionizer. In: Journal of Microelectromechanical Systems. 2008 ; Vol. 17, No. 1. pp. 115-123.
@article{98a5e5ca797f46c2a78468714ae603b0,
title = "Design, fabrication, and testing of a microfabricated Corona Ionizer",
abstract = "A microfabricated corona ionizer is developed for miniaturized air particle monitoring instruments. Peek's breakdown criterion and Warburg's Law are used to calculate the required electrode geometries. Single mask electroplating is employed as the microfabrication technique. Devices with discharge pin radius of 20 μm and interelectrode spacing of 1.4, 1.8, 2.0, and 2.2 mm are fabricated and tested. The inception voltages in the range from 1.4 to 2 kV are achieved without substrate breakdown. It consumes relatively low power at 150 mW or less. Corona charging currents of more than 50 μA were achieved before streamers break down. The experimental results also demonstrated the nondependency of the inception voltage on the pin's radius of curvature so long as it is significantly smaller than the interelectrode spacing. There is only one value of inception voltage for each given interelectrode spacing (1.4 to 2.2 mm) with the pin's radius of curvature ranging from 10 to 200 μm. This implies that strict dimensional control is not required in the microfabrication process in order to ensure the same inception voltage for each device.",
keywords = "Avalanche, Corona, Discharge, Electroplating, Inception, Ionizer, Microfabricated, Plasma, Streamer",
author = "Beelee Chua and Wexler, {Anthony S.} and Tien, {Norman C.} and Niemeier, {Debbie A.} and Holmen, {Britt A.}",
year = "2008",
month = "2",
doi = "10.1109/JMEMS.2007.909515",
language = "English (US)",
volume = "17",
pages = "115--123",
journal = "Journal of Microelectromechanical Systems",
issn = "1057-7157",
publisher = "Institute of Electrical and Electronics Engineers Inc.",
number = "1",

}

TY - JOUR

T1 - Design, fabrication, and testing of a microfabricated Corona Ionizer

AU - Chua, Beelee

AU - Wexler, Anthony S.

AU - Tien, Norman C.

AU - Niemeier, Debbie A.

AU - Holmen, Britt A.

PY - 2008/2

Y1 - 2008/2

N2 - A microfabricated corona ionizer is developed for miniaturized air particle monitoring instruments. Peek's breakdown criterion and Warburg's Law are used to calculate the required electrode geometries. Single mask electroplating is employed as the microfabrication technique. Devices with discharge pin radius of 20 μm and interelectrode spacing of 1.4, 1.8, 2.0, and 2.2 mm are fabricated and tested. The inception voltages in the range from 1.4 to 2 kV are achieved without substrate breakdown. It consumes relatively low power at 150 mW or less. Corona charging currents of more than 50 μA were achieved before streamers break down. The experimental results also demonstrated the nondependency of the inception voltage on the pin's radius of curvature so long as it is significantly smaller than the interelectrode spacing. There is only one value of inception voltage for each given interelectrode spacing (1.4 to 2.2 mm) with the pin's radius of curvature ranging from 10 to 200 μm. This implies that strict dimensional control is not required in the microfabrication process in order to ensure the same inception voltage for each device.

AB - A microfabricated corona ionizer is developed for miniaturized air particle monitoring instruments. Peek's breakdown criterion and Warburg's Law are used to calculate the required electrode geometries. Single mask electroplating is employed as the microfabrication technique. Devices with discharge pin radius of 20 μm and interelectrode spacing of 1.4, 1.8, 2.0, and 2.2 mm are fabricated and tested. The inception voltages in the range from 1.4 to 2 kV are achieved without substrate breakdown. It consumes relatively low power at 150 mW or less. Corona charging currents of more than 50 μA were achieved before streamers break down. The experimental results also demonstrated the nondependency of the inception voltage on the pin's radius of curvature so long as it is significantly smaller than the interelectrode spacing. There is only one value of inception voltage for each given interelectrode spacing (1.4 to 2.2 mm) with the pin's radius of curvature ranging from 10 to 200 μm. This implies that strict dimensional control is not required in the microfabrication process in order to ensure the same inception voltage for each device.

KW - Avalanche

KW - Corona

KW - Discharge

KW - Electroplating

KW - Inception

KW - Ionizer

KW - Microfabricated

KW - Plasma

KW - Streamer

UR - http://www.scopus.com/inward/record.url?scp=40449096694&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=40449096694&partnerID=8YFLogxK

U2 - 10.1109/JMEMS.2007.909515

DO - 10.1109/JMEMS.2007.909515

M3 - Article

AN - SCOPUS:40449096694

VL - 17

SP - 115

EP - 123

JO - Journal of Microelectromechanical Systems

JF - Journal of Microelectromechanical Systems

SN - 1057-7157

IS - 1

ER -