Comparison of four depth-encoding PET detector modules with wavelength shifting (WLS) and optical fiber read-out

Huini Du, Yongfeng Yang, Simon R Cherry

Research output: Contribution to journalArticle

15 Citations (Scopus)

Abstract

We propose detectors for a laboratory positron emission tomography scanner specific for mouse imaging that utilizes fewer detectors and channels of electronics compared with existing designs. The detectors are based on lutetium oxyorthosilicate arrays, read out by orthogonal optical fibers placed on the top and bottom of the arrays. Depth of interaction (DOI) information is obtained from the ratio of the signals at either end of the array. Four different detector modules were evaluated, using different reflector materials and two types of optical fibers (wavelength shifting (WLS) fibers and clear optical fibers). The modules were compared in terms of flood histograms, energy resolution, DOI resolution and timing resolution. Energy resolution for single crystals at one irradiation depth was around 65% full-width half-maximum (FWHM). A DOI resolution of ∼6 mm was obtained for the modules. Timing resolution was in the range of 5.1-7.8 ns. An array assembled in the laboratory and coupled with WLS fibers had the best DOI resolution; the same array with clear fibers had the best timing resolution and a commercially manufactured array and coupled with WLS fibers had the best energy resolution.

Original languageEnglish (US)
Pages (from-to)1829-1842
Number of pages14
JournalPhysics in Medicine and Biology
Volume53
Issue number7
DOIs
StatePublished - Apr 7 2008

Fingerprint

Optical Fibers
Optical fibers
coding
modules
optical fibers
Detectors
Wavelength
Fibers
detectors
wavelengths
Lutetium
Positron-Emission Tomography
Positron emission tomography
fibers
time measurement
Electronic equipment
Irradiation
Single crystals
Imaging techniques
interactions

ASJC Scopus subject areas

  • Biomedical Engineering
  • Physics and Astronomy (miscellaneous)
  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

Comparison of four depth-encoding PET detector modules with wavelength shifting (WLS) and optical fiber read-out. / Du, Huini; Yang, Yongfeng; Cherry, Simon R.

In: Physics in Medicine and Biology, Vol. 53, No. 7, 07.04.2008, p. 1829-1842.

Research output: Contribution to journalArticle

@article{949e09a8d00c4573a4170239cdea2b15,
title = "Comparison of four depth-encoding PET detector modules with wavelength shifting (WLS) and optical fiber read-out",
abstract = "We propose detectors for a laboratory positron emission tomography scanner specific for mouse imaging that utilizes fewer detectors and channels of electronics compared with existing designs. The detectors are based on lutetium oxyorthosilicate arrays, read out by orthogonal optical fibers placed on the top and bottom of the arrays. Depth of interaction (DOI) information is obtained from the ratio of the signals at either end of the array. Four different detector modules were evaluated, using different reflector materials and two types of optical fibers (wavelength shifting (WLS) fibers and clear optical fibers). The modules were compared in terms of flood histograms, energy resolution, DOI resolution and timing resolution. Energy resolution for single crystals at one irradiation depth was around 65{\%} full-width half-maximum (FWHM). A DOI resolution of ∼6 mm was obtained for the modules. Timing resolution was in the range of 5.1-7.8 ns. An array assembled in the laboratory and coupled with WLS fibers had the best DOI resolution; the same array with clear fibers had the best timing resolution and a commercially manufactured array and coupled with WLS fibers had the best energy resolution.",
author = "Huini Du and Yongfeng Yang and Cherry, {Simon R}",
year = "2008",
month = "4",
day = "7",
doi = "10.1088/0031-9155/53/7/002",
language = "English (US)",
volume = "53",
pages = "1829--1842",
journal = "Physics in Medicine and Biology",
issn = "0031-9155",
publisher = "IOP Publishing Ltd.",
number = "7",

}

TY - JOUR

T1 - Comparison of four depth-encoding PET detector modules with wavelength shifting (WLS) and optical fiber read-out

AU - Du, Huini

AU - Yang, Yongfeng

AU - Cherry, Simon R

PY - 2008/4/7

Y1 - 2008/4/7

N2 - We propose detectors for a laboratory positron emission tomography scanner specific for mouse imaging that utilizes fewer detectors and channels of electronics compared with existing designs. The detectors are based on lutetium oxyorthosilicate arrays, read out by orthogonal optical fibers placed on the top and bottom of the arrays. Depth of interaction (DOI) information is obtained from the ratio of the signals at either end of the array. Four different detector modules were evaluated, using different reflector materials and two types of optical fibers (wavelength shifting (WLS) fibers and clear optical fibers). The modules were compared in terms of flood histograms, energy resolution, DOI resolution and timing resolution. Energy resolution for single crystals at one irradiation depth was around 65% full-width half-maximum (FWHM). A DOI resolution of ∼6 mm was obtained for the modules. Timing resolution was in the range of 5.1-7.8 ns. An array assembled in the laboratory and coupled with WLS fibers had the best DOI resolution; the same array with clear fibers had the best timing resolution and a commercially manufactured array and coupled with WLS fibers had the best energy resolution.

AB - We propose detectors for a laboratory positron emission tomography scanner specific for mouse imaging that utilizes fewer detectors and channels of electronics compared with existing designs. The detectors are based on lutetium oxyorthosilicate arrays, read out by orthogonal optical fibers placed on the top and bottom of the arrays. Depth of interaction (DOI) information is obtained from the ratio of the signals at either end of the array. Four different detector modules were evaluated, using different reflector materials and two types of optical fibers (wavelength shifting (WLS) fibers and clear optical fibers). The modules were compared in terms of flood histograms, energy resolution, DOI resolution and timing resolution. Energy resolution for single crystals at one irradiation depth was around 65% full-width half-maximum (FWHM). A DOI resolution of ∼6 mm was obtained for the modules. Timing resolution was in the range of 5.1-7.8 ns. An array assembled in the laboratory and coupled with WLS fibers had the best DOI resolution; the same array with clear fibers had the best timing resolution and a commercially manufactured array and coupled with WLS fibers had the best energy resolution.

UR - http://www.scopus.com/inward/record.url?scp=42449159363&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=42449159363&partnerID=8YFLogxK

U2 - 10.1088/0031-9155/53/7/002

DO - 10.1088/0031-9155/53/7/002

M3 - Article

VL - 53

SP - 1829

EP - 1842

JO - Physics in Medicine and Biology

JF - Physics in Medicine and Biology

SN - 0031-9155

IS - 7

ER -