Abstract
Bioimpedance analysis (BIA) is accepted for the assessment of total-body water (TBW), intracellular fluid (ICF) and extracellular fluid (ECF). We aimed to compare precision and accuracy of single and multi-frequency-BIA to direct estimation methods (DEMs) of TBW, ECF, and ICF in hemodialysis patients. Linear regression analysis of volume estimates in 49 patients by single- and multi-frequency-BIA correlated significantly with DEMs. Bland-Altman analysis (BAA) found systemic bias for ECF single-frequency-BIA vs. ECF-DEMs. No other systematic biases were found. Proportional errors were found by BAA of ICF and ECF assessments with single- and multi-frequency bioimpedance spectroscopy compared to the DEMs. Comparisons of indirect methods (IEMs) to DEMs showed no significant differences and proportional errors. Root mean-squared-error analysis suggested slightly better accuracy and precision of ICF single-frequency-BIA vs. DEMs over ICF multi-frequency-BIA and IEMs to DEMs, and slightly better performance for ECF multi-frequency-BIA over both respective other methods. Compared to DEMs, there is slightly better accuracy for ECF multi- over single-frequency-BIA and ICF single- over multi-frequency-BIA. However the margin of differences between direct and indirect methods suggests that none of the analyzed methods served as a true "gold standard", because indirect methods are almost equally precise compared to DEMs.
Original language | English (US) |
---|---|
Pages (from-to) | 898-908 |
Number of pages | 11 |
Journal | Kidney International |
Volume | 85 |
Issue number | 4 |
DOIs | |
State | Published - 2014 |
Keywords
- Chronic hemodialysis
- Hemodialysis
- Nutrition
- Water-electrolyte balance
ASJC Scopus subject areas
- Nephrology
- Medicine(all)