Augmented tendon achilles repair using a tissue reinforcement scaffold: A biomechanical study

Eric Giza, Lauren Frizzell, Randall Farac, Joel Williams, Sunny Kim

Research output: Contribution to journalArticle

15 Citations (Scopus)

Abstract

Background: Missed or chronic Achilles tendon ruptures may have muscle atrophy and tendon retraction resulting in a defect that must be augmented with endogenous or exogenous materials. The Artelon® Tissue Reinforcement (ATR) scaffold is a readily available synthetic degradable poly(urethane urea) material used to augment tendon repair. The objective of this study was to compare human cadaveric Achilles tendon repairs with and without ATR. Materials and Methods: Eighteen fresh frozen human cadaver limbs were dissected and the tendon transected 2 cm proximal to the calcaneal insertion. The control group of nine specimens was repaired with sutures, while the experimental group was repaired with sutures and reinforced with a tubularized patch of ATR. Specimens were tested for ultimate load to failure in an Instron machine after preloading to 10 N followed by cyclic loading for 20 cycles from 2 to 30 N. Results: The ultimate load to failure in the control group was a mean of 248.1 N ± 19.6 (202 to 293 at 95% CI) versus 370.4 N ± 25.2 (312 to 428 at 95% CI) in the ATR group. The ultimate load to failure was 370.4 ± 25.2 N (312 to 428 at 95% CI) and 248.1 ± 19.6 N (202 to 293 at 95% CI) in the experimental and control groups, respectively (p = 0.0015). Creep of the ATR augmented group was 2.0 ± 0.5 mm, compared to 3.1 ± 1.1 mm for the control group (p = 0.026). Conclusion: ATR provided a statistically significant improvement in load to failure when compared to control specimens in a cadaver model. Clinical Relevance: This finding may allow for development of more aggressive rehabilitation techniques following chronic Achilles tendon repairs.

Original languageEnglish (US)
Pages (from-to)545-549
Number of pages5
JournalFoot and Ankle International
Volume32
Issue number5
DOIs
StatePublished - May 2011

Fingerprint

Tissue Scaffolds
Achilles Tendon
Tendons
Control Groups
Cadaver
Sutures
Muscular Atrophy
Urethane
Urea
polyetherurethane urea
Rupture
Rehabilitation
Extremities

Keywords

  • Achilles tendon
  • Repair
  • Rupture
  • Tissue reinforcement

ASJC Scopus subject areas

  • Surgery
  • Orthopedics and Sports Medicine

Cite this

Augmented tendon achilles repair using a tissue reinforcement scaffold : A biomechanical study. / Giza, Eric; Frizzell, Lauren; Farac, Randall; Williams, Joel; Kim, Sunny.

In: Foot and Ankle International, Vol. 32, No. 5, 05.2011, p. 545-549.

Research output: Contribution to journalArticle

Giza, Eric ; Frizzell, Lauren ; Farac, Randall ; Williams, Joel ; Kim, Sunny. / Augmented tendon achilles repair using a tissue reinforcement scaffold : A biomechanical study. In: Foot and Ankle International. 2011 ; Vol. 32, No. 5. pp. 545-549.
@article{df8d2f896a0d41f09bca9a711f26bbe5,
title = "Augmented tendon achilles repair using a tissue reinforcement scaffold: A biomechanical study",
abstract = "Background: Missed or chronic Achilles tendon ruptures may have muscle atrophy and tendon retraction resulting in a defect that must be augmented with endogenous or exogenous materials. The Artelon{\circledR} Tissue Reinforcement (ATR) scaffold is a readily available synthetic degradable poly(urethane urea) material used to augment tendon repair. The objective of this study was to compare human cadaveric Achilles tendon repairs with and without ATR. Materials and Methods: Eighteen fresh frozen human cadaver limbs were dissected and the tendon transected 2 cm proximal to the calcaneal insertion. The control group of nine specimens was repaired with sutures, while the experimental group was repaired with sutures and reinforced with a tubularized patch of ATR. Specimens were tested for ultimate load to failure in an Instron machine after preloading to 10 N followed by cyclic loading for 20 cycles from 2 to 30 N. Results: The ultimate load to failure in the control group was a mean of 248.1 N ± 19.6 (202 to 293 at 95{\%} CI) versus 370.4 N ± 25.2 (312 to 428 at 95{\%} CI) in the ATR group. The ultimate load to failure was 370.4 ± 25.2 N (312 to 428 at 95{\%} CI) and 248.1 ± 19.6 N (202 to 293 at 95{\%} CI) in the experimental and control groups, respectively (p = 0.0015). Creep of the ATR augmented group was 2.0 ± 0.5 mm, compared to 3.1 ± 1.1 mm for the control group (p = 0.026). Conclusion: ATR provided a statistically significant improvement in load to failure when compared to control specimens in a cadaver model. Clinical Relevance: This finding may allow for development of more aggressive rehabilitation techniques following chronic Achilles tendon repairs.",
keywords = "Achilles tendon, Repair, Rupture, Tissue reinforcement",
author = "Eric Giza and Lauren Frizzell and Randall Farac and Joel Williams and Sunny Kim",
year = "2011",
month = "5",
doi = "10.3113/FAI.2011.0545",
language = "English (US)",
volume = "32",
pages = "545--549",
journal = "Foot and Ankle International",
issn = "1071-1007",
publisher = "AOFAS - American Orthopaedic Foot and Ankle Society",
number = "5",

}

TY - JOUR

T1 - Augmented tendon achilles repair using a tissue reinforcement scaffold

T2 - A biomechanical study

AU - Giza, Eric

AU - Frizzell, Lauren

AU - Farac, Randall

AU - Williams, Joel

AU - Kim, Sunny

PY - 2011/5

Y1 - 2011/5

N2 - Background: Missed or chronic Achilles tendon ruptures may have muscle atrophy and tendon retraction resulting in a defect that must be augmented with endogenous or exogenous materials. The Artelon® Tissue Reinforcement (ATR) scaffold is a readily available synthetic degradable poly(urethane urea) material used to augment tendon repair. The objective of this study was to compare human cadaveric Achilles tendon repairs with and without ATR. Materials and Methods: Eighteen fresh frozen human cadaver limbs were dissected and the tendon transected 2 cm proximal to the calcaneal insertion. The control group of nine specimens was repaired with sutures, while the experimental group was repaired with sutures and reinforced with a tubularized patch of ATR. Specimens were tested for ultimate load to failure in an Instron machine after preloading to 10 N followed by cyclic loading for 20 cycles from 2 to 30 N. Results: The ultimate load to failure in the control group was a mean of 248.1 N ± 19.6 (202 to 293 at 95% CI) versus 370.4 N ± 25.2 (312 to 428 at 95% CI) in the ATR group. The ultimate load to failure was 370.4 ± 25.2 N (312 to 428 at 95% CI) and 248.1 ± 19.6 N (202 to 293 at 95% CI) in the experimental and control groups, respectively (p = 0.0015). Creep of the ATR augmented group was 2.0 ± 0.5 mm, compared to 3.1 ± 1.1 mm for the control group (p = 0.026). Conclusion: ATR provided a statistically significant improvement in load to failure when compared to control specimens in a cadaver model. Clinical Relevance: This finding may allow for development of more aggressive rehabilitation techniques following chronic Achilles tendon repairs.

AB - Background: Missed or chronic Achilles tendon ruptures may have muscle atrophy and tendon retraction resulting in a defect that must be augmented with endogenous or exogenous materials. The Artelon® Tissue Reinforcement (ATR) scaffold is a readily available synthetic degradable poly(urethane urea) material used to augment tendon repair. The objective of this study was to compare human cadaveric Achilles tendon repairs with and without ATR. Materials and Methods: Eighteen fresh frozen human cadaver limbs were dissected and the tendon transected 2 cm proximal to the calcaneal insertion. The control group of nine specimens was repaired with sutures, while the experimental group was repaired with sutures and reinforced with a tubularized patch of ATR. Specimens were tested for ultimate load to failure in an Instron machine after preloading to 10 N followed by cyclic loading for 20 cycles from 2 to 30 N. Results: The ultimate load to failure in the control group was a mean of 248.1 N ± 19.6 (202 to 293 at 95% CI) versus 370.4 N ± 25.2 (312 to 428 at 95% CI) in the ATR group. The ultimate load to failure was 370.4 ± 25.2 N (312 to 428 at 95% CI) and 248.1 ± 19.6 N (202 to 293 at 95% CI) in the experimental and control groups, respectively (p = 0.0015). Creep of the ATR augmented group was 2.0 ± 0.5 mm, compared to 3.1 ± 1.1 mm for the control group (p = 0.026). Conclusion: ATR provided a statistically significant improvement in load to failure when compared to control specimens in a cadaver model. Clinical Relevance: This finding may allow for development of more aggressive rehabilitation techniques following chronic Achilles tendon repairs.

KW - Achilles tendon

KW - Repair

KW - Rupture

KW - Tissue reinforcement

UR - http://www.scopus.com/inward/record.url?scp=79955734622&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=79955734622&partnerID=8YFLogxK

U2 - 10.3113/FAI.2011.0545

DO - 10.3113/FAI.2011.0545

M3 - Article

C2 - 21733465

AN - SCOPUS:79955734622

VL - 32

SP - 545

EP - 549

JO - Foot and Ankle International

JF - Foot and Ankle International

SN - 1071-1007

IS - 5

ER -