Applications of specialized coils for high-resolution MRI on a whole-body scanner

Jürgen Mäurer, Herman Requardt, Bernhard Sander, Friedrich D Knollmann, Arne Jörn Lemke, Thomas J. Vogl, Roland Felix

Research output: Contribution to journalArticle

8 Citations (Scopus)

Abstract

To investigate the application of a mini coil surface system for high- resolution MRI, 60 volunteers were examined in a 1.5-T whole-body scanner. Two replaceable probe heads were available: a circular 2.5-cm coil and a quadratic 5-cm coil, both of which were placed directly on the skin. The skin layers, Achilles tendon and finger joints were examined with the 2.5 cm coil and a FOV of 25 x 25 mm2. A matrix of 256 x 256 pixels resulted in a pixel size of 0.098 x 0.098 mm2. For imaging of the carpal tunnel, the 5 cm coil was used in transverse orientation. The FOV was 50 x 50 mm2 so that a matrix of 256 x 256 pixels led to a pixel size of 0.195 x 0.195 mm2. The resulting spatial resolution permitted visualization of the epidermis, dermis and subcutis, resulting in clear definition of anatomical detail of the musculoskeletal system. MRI measurement of skin-layer thickness did not correlate with histometric data (p < 0.05). This discrepancy was due in part to shrinkage of the tumor specimen on histologic preparation. Other causes include the motion artifacts and the limited accuracy of determining thickness on the MRI display unit.

Original languageEnglish (US)
Pages (from-to)27-33
Number of pages7
JournalMagnetic Resonance Materials in Physics, Biology, and Medicine
Volume4
Issue number1
DOIs
StatePublished - 1996
Externally publishedYes

Fingerprint

Skin
Finger Joint
Musculoskeletal System
Achilles Tendon
Dermis
Wrist
Epidermis
Artifacts
Volunteers
Head
Neoplasms

Keywords

  • Achilles tendon
  • finger
  • magnetic resonance imaging
  • MR-imaging technology
  • skin
  • surface coils
  • tissue characterization
  • wrist

ASJC Scopus subject areas

  • Biophysics
  • Genetics

Cite this

Applications of specialized coils for high-resolution MRI on a whole-body scanner. / Mäurer, Jürgen; Requardt, Herman; Sander, Bernhard; Knollmann, Friedrich D; Lemke, Arne Jörn; Vogl, Thomas J.; Felix, Roland.

In: Magnetic Resonance Materials in Physics, Biology, and Medicine, Vol. 4, No. 1, 1996, p. 27-33.

Research output: Contribution to journalArticle

Mäurer, Jürgen ; Requardt, Herman ; Sander, Bernhard ; Knollmann, Friedrich D ; Lemke, Arne Jörn ; Vogl, Thomas J. ; Felix, Roland. / Applications of specialized coils for high-resolution MRI on a whole-body scanner. In: Magnetic Resonance Materials in Physics, Biology, and Medicine. 1996 ; Vol. 4, No. 1. pp. 27-33.
@article{b1d3f6deebbb46d08466498b49049eae,
title = "Applications of specialized coils for high-resolution MRI on a whole-body scanner",
abstract = "To investigate the application of a mini coil surface system for high- resolution MRI, 60 volunteers were examined in a 1.5-T whole-body scanner. Two replaceable probe heads were available: a circular 2.5-cm coil and a quadratic 5-cm coil, both of which were placed directly on the skin. The skin layers, Achilles tendon and finger joints were examined with the 2.5 cm coil and a FOV of 25 x 25 mm2. A matrix of 256 x 256 pixels resulted in a pixel size of 0.098 x 0.098 mm2. For imaging of the carpal tunnel, the 5 cm coil was used in transverse orientation. The FOV was 50 x 50 mm2 so that a matrix of 256 x 256 pixels led to a pixel size of 0.195 x 0.195 mm2. The resulting spatial resolution permitted visualization of the epidermis, dermis and subcutis, resulting in clear definition of anatomical detail of the musculoskeletal system. MRI measurement of skin-layer thickness did not correlate with histometric data (p < 0.05). This discrepancy was due in part to shrinkage of the tumor specimen on histologic preparation. Other causes include the motion artifacts and the limited accuracy of determining thickness on the MRI display unit.",
keywords = "Achilles tendon, finger, magnetic resonance imaging, MR-imaging technology, skin, surface coils, tissue characterization, wrist",
author = "J{\"u}rgen M{\"a}urer and Herman Requardt and Bernhard Sander and Knollmann, {Friedrich D} and Lemke, {Arne J{\"o}rn} and Vogl, {Thomas J.} and Roland Felix",
year = "1996",
doi = "10.1007/BF01759777",
language = "English (US)",
volume = "4",
pages = "27--33",
journal = "Magnetic Resonance Materials in Physics, Biology, and Medicine",
issn = "0968-5243",
publisher = "Springer Verlag",
number = "1",

}

TY - JOUR

T1 - Applications of specialized coils for high-resolution MRI on a whole-body scanner

AU - Mäurer, Jürgen

AU - Requardt, Herman

AU - Sander, Bernhard

AU - Knollmann, Friedrich D

AU - Lemke, Arne Jörn

AU - Vogl, Thomas J.

AU - Felix, Roland

PY - 1996

Y1 - 1996

N2 - To investigate the application of a mini coil surface system for high- resolution MRI, 60 volunteers were examined in a 1.5-T whole-body scanner. Two replaceable probe heads were available: a circular 2.5-cm coil and a quadratic 5-cm coil, both of which were placed directly on the skin. The skin layers, Achilles tendon and finger joints were examined with the 2.5 cm coil and a FOV of 25 x 25 mm2. A matrix of 256 x 256 pixels resulted in a pixel size of 0.098 x 0.098 mm2. For imaging of the carpal tunnel, the 5 cm coil was used in transverse orientation. The FOV was 50 x 50 mm2 so that a matrix of 256 x 256 pixels led to a pixel size of 0.195 x 0.195 mm2. The resulting spatial resolution permitted visualization of the epidermis, dermis and subcutis, resulting in clear definition of anatomical detail of the musculoskeletal system. MRI measurement of skin-layer thickness did not correlate with histometric data (p < 0.05). This discrepancy was due in part to shrinkage of the tumor specimen on histologic preparation. Other causes include the motion artifacts and the limited accuracy of determining thickness on the MRI display unit.

AB - To investigate the application of a mini coil surface system for high- resolution MRI, 60 volunteers were examined in a 1.5-T whole-body scanner. Two replaceable probe heads were available: a circular 2.5-cm coil and a quadratic 5-cm coil, both of which were placed directly on the skin. The skin layers, Achilles tendon and finger joints were examined with the 2.5 cm coil and a FOV of 25 x 25 mm2. A matrix of 256 x 256 pixels resulted in a pixel size of 0.098 x 0.098 mm2. For imaging of the carpal tunnel, the 5 cm coil was used in transverse orientation. The FOV was 50 x 50 mm2 so that a matrix of 256 x 256 pixels led to a pixel size of 0.195 x 0.195 mm2. The resulting spatial resolution permitted visualization of the epidermis, dermis and subcutis, resulting in clear definition of anatomical detail of the musculoskeletal system. MRI measurement of skin-layer thickness did not correlate with histometric data (p < 0.05). This discrepancy was due in part to shrinkage of the tumor specimen on histologic preparation. Other causes include the motion artifacts and the limited accuracy of determining thickness on the MRI display unit.

KW - Achilles tendon

KW - finger

KW - magnetic resonance imaging

KW - MR-imaging technology

KW - skin

KW - surface coils

KW - tissue characterization

KW - wrist

UR - http://www.scopus.com/inward/record.url?scp=0029952587&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0029952587&partnerID=8YFLogxK

U2 - 10.1007/BF01759777

DO - 10.1007/BF01759777

M3 - Article

VL - 4

SP - 27

EP - 33

JO - Magnetic Resonance Materials in Physics, Biology, and Medicine

JF - Magnetic Resonance Materials in Physics, Biology, and Medicine

SN - 0968-5243

IS - 1

ER -