A study of the timing properties of position-sensitive avalanche photodiodes

Yibao Wu, Thomas S C Ng, Yongfeng Yang, Kanai Shah, Richard Farrell, Simon R Cherry

Research output: Contribution to journalArticle

15 Citations (Scopus)

Abstract

In this paper, we study position-dependent timing shifts and timing resolution in position sensitive avalanche photodiodes (PSAPDs) and their effects on the coincidence window used in positron emission tomography (PET) systems using these devices. There is a delay in PSAPD signals that increases as the excitation position moves from the corner to the center of the device and the timing resolution concurrently worsens. The difference in timing between the center and the corner can be up to 30.7 ns for a 14 × 14 mm2 area PSAPD. This means that a PSAPD-based PET system could require a very wide coincidence timing window (>60 ns) if this effect is not corrected, although the individual crystal pairs still have full-width half-maximum (FWHM) timing resolutions better than 7.4 ns. In addition to characterizing the timing properties of PSAPDs, two correction methods were developed and applied to data from a pair of PSAPD detectors. These two timing offset corrections reduced the timing shift of a crystal pair from 52.4 ns to 9.7 ns or 1.3 ns, improved the FWHM timing resolution of the detector pair from 24.6 ns to 9.5 ns or 6.0 ns and reduced the timing window (sufficient to cover at least twice the FWHM for all crystal pairs) from 65.1 ns to 22.0 ns or 15.2 ns, respectively. A two-step timing alignment method is proposed for a PET system consisting of multiple PSAPDs. Lastly, the effect of PSAPD size on the timing performance was also evaluated.

Original languageEnglish (US)
Pages (from-to)5155-5172
Number of pages18
JournalPhysics in Medicine and Biology
Volume54
Issue number17
DOIs
StatePublished - 2009

Fingerprint

Avalanches
Positron-Emission Tomography
Equipment and Supplies

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

A study of the timing properties of position-sensitive avalanche photodiodes. / Wu, Yibao; Ng, Thomas S C; Yang, Yongfeng; Shah, Kanai; Farrell, Richard; Cherry, Simon R.

In: Physics in Medicine and Biology, Vol. 54, No. 17, 2009, p. 5155-5172.

Research output: Contribution to journalArticle

Wu, Yibao ; Ng, Thomas S C ; Yang, Yongfeng ; Shah, Kanai ; Farrell, Richard ; Cherry, Simon R. / A study of the timing properties of position-sensitive avalanche photodiodes. In: Physics in Medicine and Biology. 2009 ; Vol. 54, No. 17. pp. 5155-5172.
@article{60677a0463164e5080e0ce4d29139aa6,
title = "A study of the timing properties of position-sensitive avalanche photodiodes",
abstract = "In this paper, we study position-dependent timing shifts and timing resolution in position sensitive avalanche photodiodes (PSAPDs) and their effects on the coincidence window used in positron emission tomography (PET) systems using these devices. There is a delay in PSAPD signals that increases as the excitation position moves from the corner to the center of the device and the timing resolution concurrently worsens. The difference in timing between the center and the corner can be up to 30.7 ns for a 14 × 14 mm2 area PSAPD. This means that a PSAPD-based PET system could require a very wide coincidence timing window (>60 ns) if this effect is not corrected, although the individual crystal pairs still have full-width half-maximum (FWHM) timing resolutions better than 7.4 ns. In addition to characterizing the timing properties of PSAPDs, two correction methods were developed and applied to data from a pair of PSAPD detectors. These two timing offset corrections reduced the timing shift of a crystal pair from 52.4 ns to 9.7 ns or 1.3 ns, improved the FWHM timing resolution of the detector pair from 24.6 ns to 9.5 ns or 6.0 ns and reduced the timing window (sufficient to cover at least twice the FWHM for all crystal pairs) from 65.1 ns to 22.0 ns or 15.2 ns, respectively. A two-step timing alignment method is proposed for a PET system consisting of multiple PSAPDs. Lastly, the effect of PSAPD size on the timing performance was also evaluated.",
author = "Yibao Wu and Ng, {Thomas S C} and Yongfeng Yang and Kanai Shah and Richard Farrell and Cherry, {Simon R}",
year = "2009",
doi = "10.1088/0031-9155/54/17/006",
language = "English (US)",
volume = "54",
pages = "5155--5172",
journal = "Physics in Medicine and Biology",
issn = "0031-9155",
publisher = "IOP Publishing Ltd.",
number = "17",

}

TY - JOUR

T1 - A study of the timing properties of position-sensitive avalanche photodiodes

AU - Wu, Yibao

AU - Ng, Thomas S C

AU - Yang, Yongfeng

AU - Shah, Kanai

AU - Farrell, Richard

AU - Cherry, Simon R

PY - 2009

Y1 - 2009

N2 - In this paper, we study position-dependent timing shifts and timing resolution in position sensitive avalanche photodiodes (PSAPDs) and their effects on the coincidence window used in positron emission tomography (PET) systems using these devices. There is a delay in PSAPD signals that increases as the excitation position moves from the corner to the center of the device and the timing resolution concurrently worsens. The difference in timing between the center and the corner can be up to 30.7 ns for a 14 × 14 mm2 area PSAPD. This means that a PSAPD-based PET system could require a very wide coincidence timing window (>60 ns) if this effect is not corrected, although the individual crystal pairs still have full-width half-maximum (FWHM) timing resolutions better than 7.4 ns. In addition to characterizing the timing properties of PSAPDs, two correction methods were developed and applied to data from a pair of PSAPD detectors. These two timing offset corrections reduced the timing shift of a crystal pair from 52.4 ns to 9.7 ns or 1.3 ns, improved the FWHM timing resolution of the detector pair from 24.6 ns to 9.5 ns or 6.0 ns and reduced the timing window (sufficient to cover at least twice the FWHM for all crystal pairs) from 65.1 ns to 22.0 ns or 15.2 ns, respectively. A two-step timing alignment method is proposed for a PET system consisting of multiple PSAPDs. Lastly, the effect of PSAPD size on the timing performance was also evaluated.

AB - In this paper, we study position-dependent timing shifts and timing resolution in position sensitive avalanche photodiodes (PSAPDs) and their effects on the coincidence window used in positron emission tomography (PET) systems using these devices. There is a delay in PSAPD signals that increases as the excitation position moves from the corner to the center of the device and the timing resolution concurrently worsens. The difference in timing between the center and the corner can be up to 30.7 ns for a 14 × 14 mm2 area PSAPD. This means that a PSAPD-based PET system could require a very wide coincidence timing window (>60 ns) if this effect is not corrected, although the individual crystal pairs still have full-width half-maximum (FWHM) timing resolutions better than 7.4 ns. In addition to characterizing the timing properties of PSAPDs, two correction methods were developed and applied to data from a pair of PSAPD detectors. These two timing offset corrections reduced the timing shift of a crystal pair from 52.4 ns to 9.7 ns or 1.3 ns, improved the FWHM timing resolution of the detector pair from 24.6 ns to 9.5 ns or 6.0 ns and reduced the timing window (sufficient to cover at least twice the FWHM for all crystal pairs) from 65.1 ns to 22.0 ns or 15.2 ns, respectively. A two-step timing alignment method is proposed for a PET system consisting of multiple PSAPDs. Lastly, the effect of PSAPD size on the timing performance was also evaluated.

UR - http://www.scopus.com/inward/record.url?scp=70350514271&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=70350514271&partnerID=8YFLogxK

U2 - 10.1088/0031-9155/54/17/006

DO - 10.1088/0031-9155/54/17/006

M3 - Article

C2 - 19671971

AN - SCOPUS:70350514271

VL - 54

SP - 5155

EP - 5172

JO - Physics in Medicine and Biology

JF - Physics in Medicine and Biology

SN - 0031-9155

IS - 17

ER -